Skip to main content

Table 4 Criterion validity of the iPhone compared to the CROM

From: Reliability and criterion validity of two applications of the iPhone™ to measure cervical range of motion in healthy participants

 

iPhone (Mean ± SD)

CROM (Mean ± SD)

ICC ± [95% CI]

Pearson’s r (p-value)

Flexion

56.0° ± 8.3°

57.1° ± 8.4°

0.76 [0.55-0.88]

0.69 p < 0.001

Extension

82.2° ± 14.0°

85.5° ± 12.3°

0.58 [0.27-0.78]

0.56 p = 0.002

Right lateral flexion

45.6° ± 7.8°

44.7° ± 8.2°

0.85 [0.70-0.93]

0.80 p < 0.001

Left lateral flexion

47.5° ± 6.8°

47.2° ± 7.0°

0.70 [0.46-0.85]

0.63 p < 0.001

Right rotation

72.4° ± 14.5°

73.8° ± 8.6°

0.55 [0.23-0.76]

0.58 P < 0.01

Left rotation

75.4° ± 14.7°

74.9° ± 8.6°

0.43 [0.08-0.69]

0.38 p = 0.04