From: The design and testing of a novel mechanomyogram-driven switch controlled by small eyebrow movements
Participant | Contraction detection | Attempt rating | Effort rating | Slope of response time | ||
---|---|---|---|---|---|---|
Sensitivity | Specificity | 95% CI of slope (ms/min) | ||||
B1 | 1.000 | 1.000 | 1 | 2 | -10.25 | -1.75 |
A1 | 1.000 | 1.000 | 1 | 2 | -4.17 | -0.31 |
A2 | 1.000 | 1.000 | 1 | 2 | -9.71 | -0.11 |
A3 | 1.000 | 1.000 | 1 | 2 | -0.15 | 6.77 |
A4 | 1.000 | 0.997 | 2 | 2 | -4.19 | 4.92 |
A5 | 0.990 | 1.000 | 2 | 2 | -1.59 | 2.55 |
A6 | 1.000 | 1.000 | 2 | 3 | -5.83 | 3.57 |
A7 | 1.000 | 1.000 | 2 | 1 | -4.36 | 5.35 |
A8 | 0.990 | 1.000 | 1 | 2 | -14.64 | -5.84 |
A9 | 0.990 | 0.997 | 2 | 3 | -2.33 | 9.54 |
A10 | 1.000 | 1.000 | 1 | 2 | 5.84 | 11.69 |
Average | 0.997 ± 0.004 | 0.999 ± 0.001 | 1.45 ± 0.5 | 2.1 ± 0.5 | -4.67 ± 5.5 | 3.30 ± 5.1 |