From: Validation of a wireless dry electrode system for electroencephalography
Variable | Versus Headset | Mitsar - 201 |
---|---|---|
Platforms | Windows XP, 7, 8 | Window XP, 7, 8 |
Electrodes | Integrated - Carbon-Silicon | Single lead - 9 mm flat DIN gold |
Fixation | Headset, direct contact, paste-free | Direct contact, conductive paste |
Amplifier Power Supply | Rechargeable Li-poly (micro-USB) | Direct PC connection (USB) |
Amplifier Current Absorption | 60 mA | 100 mA |
Continuous Operation Time | 5 h | 8 h |
Channels | 5 (Fz, C3, Cz, C4, Pz) | 21 |
Reference | 1 (A1) | 2 (A1 + A2) |
Data Transmission | Bluetooth 2.1 | USB |
RF Frequency | 2.4-2.48 GHz | N/A |
RF Range | 10 m | N/A |
UART Baud Rate | 115,200 | 460,800 |
Hardware Bandwidth | 1-1 k Hz | 0.16 - 70 Hz |
Filter Type | Elliptical | IIR |
Filter Order | Multiple | Multiple |
Input Voltage Range | 0.4 – 820 μV | 1.5 - 5000 μV |
Input Impedance | 100 MΩ | Not < 200 MΩ |
Electrode Impedance | ~ 100 Ω | < 5 kΩ |
Input Referred Noise | < 0.4 μVpp | < 1.5 μVpp |
ADC Resolutions | 12-bit | 16-bit |
Sampling rate | 250/1280 sample/s | 250/500 sample/s |
Weight | 344 g | 0.9 kg |