Skip to main content

Table 2 Calculated ICCs (r-values) for all subjects (Control + Stroke), neurologically-intact only subjects (Control), and subjects with stroke (Stroke)

From: Inter-rater reliability of kinesthetic measurements with the KINARM robotic exoskeleton

Parameter Group
NO VISION Control + Stroke Control Stroke
 IDE 0.86* 0.52* 0.81*
 IDEv 0.83* 0.41 0.84*
 PLR 0.69* 0.59* 0.68
 PLRv 0.93* 0.52* 0.95*
 RL 0.95* 0.72* 0.92*
 RLv 0.94* 0.10 0.91*
 PSR 0.72* 0.72* 0.71
 PSRv 0.80* 0.30 0.86*
VISION
 IDE 0.94* 0.64* 0.95*
 IDEv 0.93* 0.37 0.94*
 PLR 0.61* 0.73* 0.44
 PLRv 0.95* 0.49 0.97*
 RL 0.97* 0.83* 0.94*
 RLv 0.96* 0.57 0.95*
 PSR 0.90* 0.71* 0.96*
 PSRv 0.77* 0.41 0.66
  1. *indicates p < 0.0063, Bonferroni corrected (n = 8, α = 0.05)