From: “Long-term stability of stimulating spiral nerve cuff electrodes on human peripheral nerves”
Subject | Implant Date | Outcome Measures | Pulse Amplitude |
---|---|---|---|
S01 | 11/8/2005 | 8 charge threshold sessions from one month to 10.4 years post-implantation. | 0.8 mA in 58% of the trials; 1.4 mA in 24% of the trials; 2.1 mA in 18% of the trials |
S07 | 8/30/2010 | 10 charge threshold sessions from one month to 5.8 years post-implantation; measured moments to analyze muscle selectivity at 6.3 years post-implantation | 1.4 mA |
S08 | 5/2/2011 | 11 charge threshold sessions from one month to 5.3 years post-implantation; measured moments to analyze muscle selectivity at 4.2 years post-implantation | 0.8 mA |
S09 | 1/16/2012 | 6 recruitment curves at 4.5 years post-implantation | Left femoral: 2.1 mA Right femoral: 18 mA Left tibial: 2.1 mA Right tibial: 1.4 mA Left fibular: 1.4 mA Right fibular: 20 mA |
S10 | 5/15/2012 | 6 recruitment curves at 4.5 years post-implantation | Left femoral: 1.4 mA Right femoral: 0.8 mA Right tibial: 2.1 mA Right fibular: 2.1 mA Left tibial: 1.4 mA Left fibular: 2.1 mA |
S11 | 5/24/2012 | 13 charge threshold sessions from three weeks to 4.6 years post-implantation. | Radial contact #1: 0.5–0.7 mA Radial contact #2: 0.4–0.7 mA Radial contact #3: 0.3–0.6 mA Radial contact #4: 0.3–0.5 mA |
S13 | 11/20/2014 | 2 recruitment curves at 2 years post-implantation | 1.5 mA |
S14 | 12/2/2014 | 2 recruitment curves at 2 years post-implantation | Left femoral: 7 mA Right femoral: 8 mA |