Table 1 Stiffness of air-structure at the inflation stages A, B, C, compared to human wrist stiffness and two medical dorsal orthosis for wrist deflection in flexion, extension, ulnar and radial deviation and the drive back

From: Design of a lightweight passive orthosis for tremor suppression

Movement

Stiffness in mNm/rad

Air-structure

Human wrist [49]

Smart Glove, IMAKa [50]

The Clutch, PRO-TEC ATHLETICSa [50]

A:40 ml

B:60 ml

C:80 ml

Flexion (φ < 0)

 Deflection

507

620

667

554

200

1500

 Drive back

220

325

408

Extension (φ > 0)

 Deflection

404

560

633

1021

200

1500

 Drive back

230

191

159

Radial deviation (ϑ > 0)

 Deflection

152

320

493

1710

300

1500

 Drive back

152

300

334

Ulnar deviation (ϑ < 0)

 Deflection

152

320

493

1245

300

1500

 Drive back

152

300

334

  1. a 1-DOF Stiffness converted from 2-DOF stiffness ellipse parameters (orthosis #4 and #6)