Table 3 Performance of the algorithm with clinical raters in classifying Non-freezers vs Freezers

From: Measuring freezing of gait during daily-life: an open-source, wearable sensors approach

Non-freezers vs freezers AUC Best Threshold Accuracy Sensitivity Specificity False positive rate False negative rate
Rater 1 vs algorithm 0.93 0.09 0.88 0.89 0.88 0.13 0.11
Rater 2 vs algorithm 0.89 0.13 0.85 0.80 0.87 0.13 0.20