Skip to main content

Table 3 Inter-and intrarater reliability for subjects with neurological movement disorders (ICC 2,1-formula).

From: Standardized voluntary force measurement in a lower extremity rehabilitation robot

 

Interrater

Intrarater

 

single measurement

average measurement

single measurement

average measurement

Joint

ICC

SEM [Nm]

CVME [%]

ICC

SEM [Nm]

CVME [%]

ICC

SEM [Nm]

CVME [%]

ICC

SEM [Nm]

CVME [%]

more affected side

            

hip flexion

0.86

6.8

11

0.90

6.2

10

0.50

11.6

26

0.79

7.6

17

hip extension

0.88

8.4

25

0.92

7.0

18

0.87

8.9

33

0.91

6.8

26

knee flexion

0.97

3.6

13

0.96

3.7

15

0.88

6.0

29

0.93

4.5

22

knee extension

0.96

4.5

9

0.85

7.9

12

0.86

8.4

22

0.86

8.2

21

less affected side

            

hip flexion

0.90

7.1

10

0.96

4.4

7

0.78

10.0

19

0.82

8.8

16

hip extension

0.66

19.3

36

0.87

11.2

21

0.81

14.6

27

0.89

10.5

20

knee flexion

0.93

5.8

16

0.95

5.0

14

0.91

6.5

20

0.96

4.1

12

knee extension

0.86

8.2

17

0.92

5.5

11

0.85

7.1

14

0.84

7.0

14

  1. Abbreviations: ICC, Intraclass correlation coefficient; SEM, standard error of measurement; CVME, coefficient of variation of method error