Skip to main content

Table 3 Concurrent validity of the VPIT with the NHPT and the BBT

From: Concurrent validity and test-retest reliability of the Virtual Peg Insertion Test to quantify upper limb function in patients with chronic stroke

Concurrent validity parameters

Spearman’s rank correlation coefficient

mean of 3 trials

All patients

Subgroup 1: without support

Subgroup 2: with support

 

(n = 31)

(n = 11)

(n = 20)

Tex [s] VPIT /

0.31

0.57

0.35

Tex [s] NHPT

0.09*

0.07*

0.13*

Tex [s] VPIT /

0.30

0.61

0.32

Tex [s] NHPT(TP1)

0.10*

0.47*

0.16*

Tex [s] VPIT /

−0.23

−0.41

−0.21

Ntc BBT

0.22*

0.22*

0.37*

Ndp VPIT /

0.21

−0.18

0.17

Ndp NHPT

0.26*

0.60*

0.49*

Ndp VPIT /

−0.12

−0.18

−0.16

Ndc BBT

0.51*

0.60*

0.51*

  1. *p-values: level of significance: p ≤ 0.05
  2. T ex [s] execution time in seconds, VPIT Virtual Peg Insertion Test, NHPT Nine Hole Peg Test, TP1 time point 1 (time stopped at the point when all 9 pegs were put in the 9 holes (without replacing them in the dish)), BBT Box and Block Test, N tc number of transported cubes, N dp number of dropped pegs, N dc number of dropped cubes, SD Standard Deviation