Skip to main content

Table 2 Reliability analysis

From: Reliability, validity, and clinical feasibility of a rapid and objective assessment of post-stroke deficits in hand proprioception

 

DL Test

DL Retest

   
 

Mean ± SD

Range

Mean ± SD

Range

ICC(2,1) [CI]

SEM

SRD

Impaired

3.08° ± 1.93°

[0.12°, 8.70°]

3.21° ± 2.22°

[0.73°, 9.97°]

0.73 [0.44,0.89]

1.07°

2.95°

Unimpaired

2.90° ± 1.94°

[0.33°, 9.53°]

2.84° ± 3.11°

[0.65°, 15.11°]

0.16 [0.00,0.56]

2.33°

6.45°

  1. Summary of the reliability analysis for the outcome measure of the robotic assessment (i.e., difference limen (DL)) of the impaired and unimpaired hand, respectively. Reported are descriptive statistics for the DL for both test and retest, reliability (ICC(2,1) and its 95% CI), standard error of measurement (SEM), and smallest real difference (SRD)