Table 2 Test-retest reliability in patients with stroke

From: Reliability, validity and discriminant ability of a robotic device for finger training in patients with subacute stroke

 

N

Test Mean (SD)

Retest Mean (SD)

ICC

95% CI

P

SEM

MDC

Lower bound

Upper bound

HandForceext (N)

115

5.7 (8.2)

5.9 (7.8)

0.977

0.967

0.984

0.365

1.2

3.4

HandForceflex (N)

115

22.1 (26.8)

21.5 (26.0)

0.979

0.969

0.985

0.467

3.8

10.6

Muscle tone (N) †

113

−12.0 (16.6)

−13.9 (17.1)

0.906

0.860

0.937

0.048

5.2

14.3

MASV1

118

1.2 (1.6)

1.2 (1.5)

0.473

0.240

0.634

0.960

1.1

3.1

MTSV1

118

0.8 (1.5)

0.9 (1.5)

0.396

0.129

0.581

0.725

1.2

3.2

MASV3

118

1.6 (1.5)

1.7 (1.5)

0.486

0.257

0.644

0.606

1.1

3.0

MTSV3

118

0.9 (1.4)

1.0 (1.5)

0.268

−0.057

0.494

0.424

1.2

3.4

  1. SD Standard Deviation; ICC Intraclass Correlation Coefficient; CI Confidence Interval. SEM Standard Error of Measurement; MDC Minimal Detectable Change. P-values in bold indicate statistical significance of the Wilcoxon signed-rank test (P lower than 0.05). † Negative values: flexor muscles tone. Positive values: extensor muscles tone