Fig. 7 | Journal of NeuroEngineering and Rehabilitation

Fig. 7

From: Chronic nerve health following implantation of femoral nerve cuff electrodes

Fig. 7

Knee moments normalized by body mass over 1 year on each of the selected knee extensor selective C-FINE contacts at 0.8 mA, 255 μs. These contacts were chosen because they were able to achieve strong knee extension moments with insignificant hip flexion. Error bars indicate standard deviation between contacts. Dashed line represents linear best-fit to mean tetanic moments. Based on linear regression t-testing, an increase in moment over 1 year is significant (p < 0.01 on the left and p < 0.02 on the right)