Ankle–foot orthosis with an oil damper versus nonarticulated ankle–foot orthosis in the gait of patients with subacute stroke: a randomized controlled trial

Abstract

Background

Gait improvement in patients with stroke has been examined in terms of use or non-use of an ankle–foot orthosis (AFO), but the effects of different kinds of AFOs remain unclear. In this study, the effect on gait of using an AFO with an oil damper (AFO-OD), which has plantarflexion stiffness without dorsiflexion resistance, was compared with a nonarticulated AFO, which has both dorsiflexion and plantarflexion stiffness, in a randomized controlled trial.

Methods

Forty-one patients (31 men, 10 women; mean age 58.4 ± 11.3 years) in the subacute phase of stroke were randomly allocated to two groups to undergo gait training for 1 h daily over 2 weeks by physiotherapists while wearing an AFO-OD or a nonarticulated AFO. A motion capture system was utilized to measure shod gait without orthosis at baseline and after training with the allocated AFO. Data analysis focused on the joint kinematics and kinetics, spatial and temporal parameters, ground reaction force, and shank-to-vertical angle. Unpaired t-test or Mann–Whitney U test was performed to clarify the difference in gait with an AFO between the two AFO groups after training, with a significance level of p = 0.05.

Results

Thirty-six patients completed the study (17 in the AFO-OD group and 19 in the nonarticulated AFO group). The ankle joint was more dorsiflexed in single stance (p = 0.008, effect size r = 0.46) and peak ankle power absorption was larger in stance (p = 0.007, r = 0.55) in the AFO-OD group compared with the nonarticulated AFO group. Peak power absorption varied among patients in the AFO-OD group. Increased dorsiflexion angles were also found at initial contact (p = 0.008, r = 1.51), pre-swing (p = 0.045, r = 0.91), and the swing phase (p = 0.045, r = 0.91) in the AFO-OD group. There was no difference in peak plantarflexion moment, ankle power generation, spatial or temporal parameters, ground reaction force, or shank-to-vertical angle between the two groups.

Conclusions

The results of this study showed that an AFO with plantarflexion stiffness but without dorsiflexion resistance produced greater improvement in ankle joint kinematics and kinetics compared with the nonarticulated AFO, but the results of peak power absorption varied greatly among patients.

Trial registration UMIN000028126, Registered 1 August 2017, https://upload.umin.ac.jp/cgi-bin/icdr/ctr_menu_form_reg.cgi?recptno=R000032197

Introduction

Ankle–foot orthoses (AFOs) are commonly used in clinical practice to improve gait in patients with stroke. Systematic reviews have shown that AFOs can improve gait in these patients by modifying the kinematics of the ankle and knee joints in the stance phase and preventing foot drop in the swing phase [1, 2]. However, the evidence for use of an AFO remains limited because of a lack of sufficient information about the effects of different kinds of AFOs on gait. Considering the effect of different kinds of AFOs on gait, the stiffness of an AFO is considered a key determinant of its functional effect. Here, stiffness is defined as the slope of the curve of moment generated by compression force or resistance versus the ankle joint angle when an AFO is deformed in plantarflexion or dorsiflexion [3, 4]. As AFO stiffness directly affects the ankle joint movement a systematic review showed that the degree of stiffness of an AFO affected ankle kinematics, suggesting that greater stiffness could generally result in a decreased peak ankle joint angle in both plantarflexion and dorsiflexion [5]. Joint kinematics affect the joint kinetics around the ankle joint. Some AFOs have been developed to assist insufficient muscle activity around the ankle joint through AFO stiffness [6, 7]. Therefore, it is important to ascertain the effects of different kinds of AFOs with different stiffness on joint kinematics and kinetics during gait.

There are three main types of passive AFOs available: non-articulated (solid-ankle or rigid), hinged (articulated), and posterior leaf spring (PLS) [8]. Non-articulated and PLS AFOs generate moment by deforming the plastic material of the AFO in both dorsi- and plantarflexion. Most articulated AFOs generate mechanical stiffness by using stoppers, springs, or other devices in plantarflexion but the joints move freely in dorsiflexion. The characteristics of the AFO in plantarflexion affect the gait in the loading response and swing phase, while in dorsiflexion they affect the gait in the mid-to-late stance phase [9]. A previous study compared gait of patients wearing a non-articulated AFO and an articulated AFO, which had a mechanical joint that stopped plantarflexion; therefore, the results included the difference in AFO characteristics both in dorsi- and plantarflexion [10]

An AFO with an oil damper (AFO-OD) was developed based on gait analysis of patients with stroke [11, 12]. The resistance generated by an oil damper slows down the rate of angular change in the ankle joint when it moves to plantarflexion in the loading response. The ankle joint of the AFO-OD moves freely to dorsiflexion. In the present randomized controlled trial, the effects of the non-articulated AFO, which has dorsi- and plantarflexion stiffness, and the AFO-OD, which has plantarflexion stiffness but not in dorsiflexion, on the gait of patients with stroke were compared. This comparison should clarify the difference between AFOs with and without dorsiflexion stiffness. Although there are differences in the characteristics of plantarflexion stiffness between the two AFOs, including the viscosity of the AFO-OD and the elasticity of the nonarticulated AFO, these might be negligible because of the small plantarflexion movement in the loading response.

The dorsiflexion stiffness of an AFO prevents excessive dorsiflexion in the mid-to-late stance phase, which offsets weakness in the plantarflexors [13, 14]; however, large stiffness in dorsiflexion impedes smooth movement of the ankle during the stance phase [1, 9, 15]. The dorsiflexion angle in mid-stance affects the progression of the center of pressure, affecting the ankle joint moment and power [16]. However, to our knowledge, no studies have compared the ankle joint kinetics between AFOs with and without dorsiflexion stiffness and how these differences affect the gait of patients with stroke. Two hypotheses were examined in this study. The first was that the dorsiflexion angle in the mid-to-terminal stance would be smaller when walking with a nonarticulated AFO than with an AFO-OD. The second was that the restricted movement of the ankle joint would induce smaller plantarflexion moment and power around the ankle joint when patients walked with a nonarticulated AFO than with an AFO-OD.

Methods

Participants

The participants were patients with subacute stroke (more than 14 and less than 180 days after onset). Patients in the subacute phase were selected for this study because the gait of patients in the chronic phase is affected largely by the use of AFOs in daily life. The participants were inpatients at a rehabilitation center in Japan (Nakaizu Rehabilitation Center) who were walking independently with or without assistive devices and had been prescribed gait training with an AFO. The participants did not use any kind of AFOs in their daily lives. Participants were assessed for eligibility by their rehabilitation physician or physical therapist from April 2018 to October 2019. Exclusion criteria were pre-existing pathology affecting the central nervous system or neuromuscular system and communication problems. In addition, patients who put the nonparetic foot behind the paretic foot (negative step length) during gait without an AFO at baseline were excluded from the analysis because this walking pattern indicates active reduction in paretic propulsion due to severe ankle plantarflexor impairment and no ankle dorsiflexion in paretic stance [17]. All participants underwent gait training under the supervision of physiotherapists. None had previously used an AFO, and all were enrolled in the study when they could walk 10 m under supervision. We calculated the sample size for performing two-way mixed analysis of variance (ANOVA) based on a previous study with peak dorsiflexion angle in the stance phase as the main outcome, assuming a variance of 5.3° and a difference of 5° between the articulated and nonarticulated AFO groups [10]. The significance level was 0.05, and the power was 0.8. Using the G-Power program (Kiel University, Kiel, Germany), the required sample size per group was calculated as 15. To account for data loss, 41 participants were recruited for this study. The study protocol was approved by the ethics committees of the International University of Health & Welfare and Nakaizu Rehabilitation Center. Written informed consent was obtained from all participants prior to their participation in this study.

Experimental AFO

The orthoses used in this study were an AFO-OD (GaitSolution; Kawamura Gishi Co. Ltd., Osaka, Japan) and a nonarticulated AFO. The AFO-OD had a mechanical ankle joint with an oil damper (Fig. 1 left) [12] that generates resistance when the ankle joint moves in plantarflexion. The amount of resistance could be changed by rotating a screw at the top of the oil damper from 1 (flexible) to 4 (rigid). In the present study, the screw was set to 3 because this level is used most frequently by patients in the subacute phase of stroke [18]. The AFO-OD allows the ankle joint to move freely in dorsiflexion without any resistance. The nonarticulated AFO used in this study had no mechanical joint and an ankle trim line behind the malleoli (Fig. 1 right). The initial ankle joint angle was set to neutral, meaning that the shank was vertical to the floor in both AFOs. Neither type of AFO were custom-made, so we prepared 6 for each group, namely, right and left AFOs in each of the three available sizes (small, medium, and large).

Fig. 1
figure 1

The ankle–foot orthoses used in this study. Left: An ankle–foot orthosis with an oil damper (AFO-OD). Right: A nonarticulated ankle–foot orthosis

The stiffness of all AFOs was measured using a goniometer, a digital force gauge (Imada Co. Ltd., Toyohashi, Japan), and a metronome. Because of differences in the inherent characteristics of plastic, which has elasticity, and the oil damper, which has viscosity, the measurement was static for the nonarticulated AFO and dynamic for the AFO-OD. The sole of the AFO was fixed on a table and the shank was pulled perpendicularly by a wire attached to the shank. The wire was pulled to 5° in both directions when obtaining measurements for the nonarticulated AFO. When recording measurements for the AFO-OD, the wire was pulled once per second and 4 times per second in plantarflexion and the peak force was recorded by the force meter. Each speed corresponds to 5°/s and 20°/s [12]. Table 1 shows the characteristics of the AFOs as the average of the right and left AFOs in each size. As shown in Table 1 dorsiflexion stiffness of the nonarticulated AFOs ranged from 1.75 to 1.92 Nm/°. In a systematic review that investigated the impact of AFO stiffness on gait, the reported range of stiffness in the included studies was 0.02–8.17 Nm/° [5]. Another study comparing four kinds of AFOs reported a stiffness range of 0.2 to 2.0 Nm/° [19]. Based on these studies, the dorsiflexion stiffness of the nonarticulated AFO used in this study may be considered moderate to relatively stiff.

Table 1 Characteristics of the AFOs used in this study

Study protocol and randomization

This single-center, non-blinded, randomized controlled, parallel group trial was registered in the University hospital Medical Information Network registry (UMIN000028216). The participants were randomly allocated to the AFO-OD group or the nonarticulated AFO group by a physiotherapist in order of participation with an allocation ratio of 1:1. First, shod gait with no AFO was measured at each participant’s self-selected walking speed along an 8-m walkway. Gait was measured using a motion capture system with 8 Vicon MX cameras (VICON Motion System, Ltd., Oxford, UK) and 6 AMTI force plates (Advanced Mechanical Technology, Inc., Phoenix, AZ). Reflective markers were attached to the participants’ bodies in accordance with the Plug-In Gait model. Additional markers were used on the medial side of the knee and ankle joints to precisely calculate the joint centers. Marker trajectories and force plate data were captured at a sampling frequency of 100 Hz. The participants wore shoes (V-step; Pacific Supply, Osaka, Japan) that were of a suitable size for each condition with and without an AFO. The use of a cane was permitted, but it had to be used in a consistent manner when walking without an AFO and with either type of AFO.

Then, the participants started gait training sessions using the allocated AFOs for 1 h daily over 2 weeks under the supervision of a team of physiotherapists. The participants used each kind of AFO just during gait training. They also participated in general physiotherapy, including range of motion exercises, balance training, and muscle training. There were no differences in the physiotherapy processes between the two AFO groups. After 2 weeks of training, gait with the allocated AFOs was measured using the same procedure.

Data analysis and outcome measures

Marker trajectories and force plate data were low-pass filtered by a second-order Butterworth filter with cutoff frequencies of 6 and 18 Hz, respectively. The gait cycle time and duration of each gait phase were defined using force plate data. The trajectories of heel markers were used if the participant walked with a cane on their nonparetic side. Given that participants were in the subacute phase of stroke, they did not show a heel rise in the paretic single stance. Therefore, the gait cycle was defined as the loading response, single stance, pre-swing, and swing phases of the paretic limb. Joint kinematics and kinetics were calculated using an inverse dynamic model. The primary outcome measure was peak dorsiflexion ankle joint angle in single stance, and the secondary outcomes were peak ankle plantarflexion moment and peak ankle power in the stance phase. The following parameters were calculated as other outcome measures: ankle joint angles in other gait phases; knee and hip joint kinematics and kinetics; spatial and temporal parameters, including velocity, step length, and duration of each gait cycle; and the anterior and posterior components of the ground reaction force. The angle of inclination of the shank segment relative to the vertical (shank-to-vertical angle) in the sagittal plane was calculated in view of its importance when evaluating gait with an AFO [20]. Step length was normalized by each participant’s body height, and ground reaction force, joint moment, and power were normalized by the participant’s body weight. Ground reaction forces as well as kinematic and kinetic data were obtained for the paretic limb. All calculations were performed using Visual 3D software version 6 (C-Motion Inc., Kingston, ON, Canada).

Statistical analysis

All gait parameters are presented as the average of at least three gait cycles for each condition. The data were assessed for normality by the Shapiro–Wilk test. For the baseline data, differences between the two AFO groups in all gait parameters without AFOs were compared using the unpaired t-test or the Mann–Whitney U test. Gait with each type of AFO after 2 weeks of gait training was compared using the unpaired t-test or Mann–Whitney U test. The correlation coefficient was also calculated between parameters that showed significant difference between groups. A p-value of < 0.05 was considered statistically significant. All statistical analyses were performed using SPSS for Windows version 25 (IBM Corp., Armonk, NY).

Results

The study flowchart is shown in Fig. 2. Recruitment started in April 2018 and was stopped in October 2019 because the required sample number was obtained. One participant in each group was withdrawn because of missing data during the study period and three participants in the AFO-OD group were excluded from the analysis because of placement of the nonparetic foot behind the paretic foot at baseline. This left data for 36 participants for analysis (AFO-OD group, n = 17; nonarticulated AFO group, n = 19). There was no significant difference between the groups in terms of age, body height, and weight or days since stroke onset at baseline (Table 2). Eleven of the 17 participants in the AFO-OD group and 15 of the 19 in the nonarticulated AFO group used a cane. No unintended effects were found during the participation.

Fig. 2
figure 2

Consolidated Standard of Reporting Trials (CONSORT) flowchart

Table 2 Patient characteristics

In the baseline comparison, no difference was found in any gait parameter between the two AFO groups (p > 0.05). The results for the comparison of joint kinematics and kinetics between the two types of AFO worn after training are shown in Table 3. For the primary outcome measure, the ankle joint was more dorsiflexed in single stance (mean difference 5.35°; 95% CI 1.50–9.19, p = 0.008, effect size r = 0.46) in the AFO-OD group than in the nonarticulated AFO group. As for the secondary outcomes, the peak ankle power absorption in stance was larger in the AFO-OD group than in the nonarticulated AFO group (mean difference 0.304 W/kg; 95% CI 0.092–0.516, p = 0.007, r = 0.55). However, there was no difference between the two AFO groups in peak ankle plantarflexion moment and peak ankle power generation in the stance phase.

Table 3 Comparison of joint kinematics and kinetics between the two AFO groups after training

As for the other outcome measures, compared with the nonarticulated AFO group, the ankle joint in the AFO-OD group was more dorsiflexed at initial contact (p = 0.008, r = 1.51), in pre-swing (p = 0.045, r = 0.91), and in the swing phase (p = 0.045, r = 0.91). The knee joint was more flexed at initial contact in the AFO-OD group than in the nonarticulated AFO group (p = 0.045, r = 0.91). The mean difference and 95% CI are not shown because these parameters were not normally distributed. There was no difference between the two AFO groups in spatial or temporal parameters, ground reaction forces, or shank-to-vertical angle (Table 4).

Table 4 Comparison of temporal and spatial factors, ground reaction forces, and shank-to-vertical angle between the two AFO groups at post-training

Figure 3 shows the mean ankle joint angle and Fig. 4 shows the ankle power in one gait cycle for each AFO group. As these figures show, the ankle joint was more dorsiflexed throughout the gait cycle and dorsiflexion in the stance phase was larger in the AFO-OD group than in the nonarticulated AFO group. Figure 4 shows that the power absorption was high with a large standard deviation in the AFO-OD group and was low with a small standard deviation in the nonarticulated AFO group.

Fig. 3
figure 3

Graph showing the average angle of the ankle joint during one gait cycle in each AFO group after training (AFO-OD group, n = 17; nonarticulated AFO group, n = 19). Bold lines indicate average values and non-bold lines denote standard deviations. AFO, ankle–foot orthosis; AFO-OD, ankle–foot orthosis with an oil damper; IC, initial contact; FO, foot off

Fig. 4
figure 4

Graph showing average ankle power during one gait cycle in each AFO group after training (AFO-OD group, n = 17; nonarticulated AFO group, n = 19). Bold lines denote average values and non-bold lines denote standard deviations. AFO, ankle–foot orthosis; AFO-OD, ankle–foot orthosis with an oil damper; IC, initial contact; FO, foot off

Figure 5 shows the relationship between the peak dorsiflexion angle in single stance and peak ankle power absorption in stance for each type of AFO. Each mark represents an individual participant (white, AFO-OD group; black, nonarticulated AFO group). Peak dorsiflexion angle and peak power absorption were large and distributed across a wide range in the AFO-OD group; there was no correlation between these parameters in the AFO-OD group (r = − 0.37, p = 0.15). Although small, these parameters were correlated with each other (r = − 0.50, p = 0.043) in the nonarticulated AFO group.

Fig. 5
figure 5

Relationship between peak angle of ankle dorsiflexion and peak power absorption at the ankle during gait in the two AFO groups. AFO, ankle–foot orthosis; AFO-OD, ankle–foot orthosis with oil damper

Discussion

We hypothesized that the peak ankle joint angle in dorsiflexion in single stance as well as the ankle plantarflexion moment and power would be smaller in the nonarticulated group. Our results partially supported these hypotheses: the peak ankle dorsiflexion angle in single stance was significantly smaller in the nonarticulated AFO group than in the AFO-OD group (1.39° vs 6.73°, p = 0.008), and the peak absorption of ankle power was also significantly smaller in the nonarticulated AFO group than in the AFO-OD group (− 0.28 W/kg vs − 0.58 W/kg, p = 0.007). However, there were no differences between the two AFO groups in ankle plantarflexion moment and peak power generation.

In terms of kinematics, the ankle joint during gait was more dorsiflexed at initial contact and in the single stance, pre-swing, and swing phases with the AFO-OD than with the nonarticulated AFO, which is consistent with previous findings [10, 21]. Systematic reviews have indicated that many previous studies focused on the peak dorsiflexion angle needed in the swing phase to keep clearance [1, 2]. The median of the peak dorsiflexion angle in the swing phase was 5.46° in the AFO-OD group and 0.05° in the nonarticulated AFO group. One of the important findings of the present study was that the peak dorsiflexion angle in the swing phase was larger with the AFO-OD than with the nonarticulated AFO.

The difference in moment generated by the two types of AFOs in plantarflexion was thought to occur in the loading response during gait. There was no difference in peak plantarflexion angle or angular velocity during the loading response between the two AFO groups. In terms of the moment in plantarflexion, the average peak plantarflexion moment generated by the nonarticulated AFO was approximately 12 Nm, with 4° plantarflexion, from − 2.48° to − 6.54°, as shown in Table 3, in loading response and 3.16 Nm/° for the AFO characteristics. We did not obtain the moment for the AFO-OD at 36°/s in plantarflexion but estimated it to be less than 10 Nm based on the results shown in Table 1. The moment to plantarflexion during loading response was similar in the two AFO groups.

The degree of stiffness in dorsiflexion provided by different types of AFO can affect the kinetics of the ankle joint in the mid-to-late stance phase [1, 9, 15]; however, to our knowledge, this had not been confirmed in any previous studies. Table 3 and Fig. 4 show that the power absorption was larger when walking with the AFO-OD than when walking with the nonarticulated AFO. Power absorption is related to eccentric contraction of the plantarflexors in mid-stance [16] and is obtained by multiplying the plantarflexion moment by the angular velocity of ankle dorsiflexion [22]. In this study, we found no significant difference in peak angular velocity of ankle dorsiflexion between the groups. Although the difference in peak plantarflexion moment in the stance phase was not significant (p = 0.054), but the main contributor to the difference in power absorption appears to be the larger plantarflexion moment in the AFO-OD group.

In the nonarticulated AFO group, the power absorption and plantarflexion moment values included both biological power and moment and AFO-generated power and moment [11, 23]. The average value of the moment for the nonarticulated AFO was approximately 2.5 Nm, calculated as 1.39° (peak dorsiflexion angle in single stance, as shown in Table 3) multiplied by 1.82 Nm/° (mean dorsiflexion stiffness of three nonarticulated AFOs, as shown in Table 1). However, the peak plantarflexion moment in stance was 47.7 Nm, calculated as 0.74 Nm/kg multiplied by average body weight 64.5 kg. The moment generated by the elasticity of the nonarticulated AFO was not large in the stance phase. This finding is in line with that in previous studies in which the AFO-generated moment during gait was estimated using a nonarticulated AFO and a testing machine [23]. In previous studies, the moment provided by an AFO was thought to be important to compensate for insufficient activity of the plantarflexors in mid-to late stance [13, 14, 24]. However, the results of the present study show that moment provided by a moderate to relatively stiff nonarticulated AFO is not large compared with the biological moment. The moment and power in the AFO-OD did not include the effect of the AFO directly because the AFO-OD did not provide any resistance in dorsiflexion. Further study is needed to determine the activity of the plantarflexors with different types of AFO.

As shown in Fig. 5, the peak power absorption was distributed in a narrow range and related to the peak angle of dorsiflexion in the nonarticulated AFO group because of the assisted dorsiflexion provided by the AFO. In the AFO-OD group, the peak power absorption was large but variable because dorsiflexion was not assisted by the AFO. The variation in ankle power absorption among patients in the AFO-OD group suggests a marked difference in plantarflexor activity between patients. The activity of the plantarflexors in stance is important in terms of stability and propulsion in normal gait [22, 25] as well as in the gait of patients with stroke [6, 7]. Therefore, an AFO-OD, which does not provide dorsiflexion resistance, would be beneficial in some patients but not others. As mentioned in previous studies [9, 26], the stiffness of the AFO should be adjusted in dorsiflexion and plantarflexion separately; however, this would be difficult for a nonarticulated AFO, where both dorsiflexion and plantarflexion are affected by the shape and thickness of the plastic materials used [27, 28].

Increased power absorption in the stance phase might affect power generation in late stance. The plantarflexors store energy and generate propulsive force in late stance [22, 25]. In this study, there was no difference in the ankle power generation between the two AFO groups. In normal gait, the ankle joint shows maximum plantarflexion in pre-swing [16] but the amount of plantarflexion was small in the AFO-OD group (Table 3). Other studies have also found a decrease in the power generated at the ankle when the AFO resisted plantarflexion [9, 29]. However, the ability to generate power can be improved by gait training using an AFO [30]. The long-term effect of gait training with an AFO on muscle activity requires further study.

This study has several limitations. The degree of stiffness of the AFO-OD in plantarflexion and of the nonarticulated AFO in plantar and dorsiflexion was the same for each participant. The fit of the AFOs, which were not custom-made, and weight differences between the two kinds of AFOs might have affected the results. The use of custom-made AFOs with optimized resistance might have had a positive effect on both AFO groups. There was a difference in the stiffness to plantarflexion between the two kinds of AFOs used in this study in terms of viscosity and elasticity. Most participants in the two groups used a cane on their nonparetic side. The use of a cane might have had an effect on the velocity and joint kinematics [31], but these participants could not walk without a cane. Finally, this study was conducted at a rehabilitation center, and thus further study in different settings will be necessary to ascertain the generalizability of the results.

Conclusion

This randomized controlled trial compared the effect of a nonarticulated AFO, which had stiffness both in dorsi- and plantarflexion, with an AFO-OD, which had stiffness in plantarflexion but not in dorsiflexion. The results of this study showed that the AFO-OD improved ankle joint kinematics and kinetics more than a nonarticulated AFO, but the results showed large variation among patients.

Availability of data and materials

The raw data were generated at Nakaizu Rehabilitation Hospital. Derived data supporting the findings of this study are available from the corresponding author S.Y. on request.

Abbreviations

AFO:

Ankle–foot orthosis

AFO-OD:

Ankle–foot orthosis with an oil damper

PLS:

Posterior leaf spring

References

  1. Daryabor A, Arazpour M, Aminian G. Effect of different designs of ankle-foot orthoses on gait in patients with stroke: a systematic review. Gait Posture. 2018;62:268–79.

    Article  Google Scholar 

  2. Tyson SF, Sadeghi-Demneh E, Nester CL. A systematic review and meta-analysis of the effect of an ankle-foot orthosis on gait biomechanics after stroke. Clin Rehabil. 2013;27:879–91.

    Article  CAS  Google Scholar 

  3. Major RE, Hewart PJ, MacDonald AM. A new structural concept in molded fixed ankle foot orthoses and comparison of the bending stiffness of four constructions. Prosthet Orthot Int. 2004;28(1):44–8.

    Article  CAS  Google Scholar 

  4. Bregman DJJ, Rozumalski A, Koops D, Groot V, Schwartz M, Harlaar J. A new method for evaluating ankle foot orthosis characteristics: BRUCE. Gait Posture. 2009;30(2):144–9.

    Article  CAS  Google Scholar 

  5. Totah D, Menon M, Jones-Hershinow C, Barton K, Gates DH. The impact of ankle-foot orthosis stiffness on gait: a systematic literature review. Gait Posture. 2019;69:101–11.

    Article  Google Scholar 

  6. Arch ES, Stanhope SJ. Passive-dynamic ankle-foot orthoses substitute for ankle strength while causing adaptive gait strategies: a feasibility study. Ann Biomed Eng. 2015;43(2):442–50.

    Article  Google Scholar 

  7. Waterval NFJ, Brehm MA, Altmann VC, Koopman FS, Den Boer JJ, Harlar J, Nollet F. Stiffness-optimized ankle-foot orthoses improve walking energy cost compared to conventional orthoses in neuromuscular disorders: a prospective uncontrolled intervention study. IEEE Trans Neural Syst Rehabil Eng. 2020;28(10):2296–304.

    Article  Google Scholar 

  8. Whittle MW, Richard J. Whittle’s gait analysis. 5th ed. Amsterdam: Elsevier; 2012.

    Google Scholar 

  9. Kobayashi T, Orendurff MS, Hunt G, Gao F, LeCursi N, Lincoln LS, Foreman KB. The effects of an articulated ankle-foot orthosis with resistance-adjustable joints on lower limb joint kinematics and kinetics during gait in individuals post-stroke. Clin Biomech. 2018;59:47–55.

    Article  Google Scholar 

  10. Gok H, Kucukdeveci A, Altinkaynak H, Yavuzer G, Ergin S. Effects of ankle-foot orthoses on hemiplaretic gait. Clin Rehabil. 2003;17:137–9.

    Article  Google Scholar 

  11. Yamamoto S, Miyazaki M, Kubota T. Quantification of the effect of the mechanical property of ankle-foot orthoses on hemiplegic gait. Gait Posture. 1993;1:27–34.

    Article  Google Scholar 

  12. Yamamoto S, Hagiwara A, Mizobe T, Yokoyama O, Yasui T. Development of an ankle-foot orthosis with an oil damper. Prosthet Orthot Int. 2005;29:209–19.

    Article  CAS  Google Scholar 

  13. Harper NG, Russel Esposito E, Wilken JM, Neptune RR. The influence of ankle-foot orthosis stiffness on walking performance in individuals with lower-limb impairment. Clin Biomech (Bristol, Avon). 2014;29:877–84.

    Article  Google Scholar 

  14. Malas BS. What reliables influence the ability of an AFO to improve function and when are they indicated? Clin Orthop Relat Res. 2011;469:1308–14.

    Article  Google Scholar 

  15. Alam M, Choudhury IA, Mamat AB. Mechanism and design analysis of articulated ankle foot orthoses for drop-foot. Sci World J. 2014;2014:867869.

    Google Scholar 

  16. Perry J, Burnfield JM. Gait analysis: normal and pathological function, SLACK; 2010.

  17. Balasubramanian CK, Bowden MG, Neptune RR, Kautz SA. Relationship between step length asymmetry and walking performance in subjects with chronic hemiparesis. Arch Phys Med Rehabil. 2007;88(1):43–9.

    Article  Google Scholar 

  18. Yamamoto S, Tanaka S, Motojima N. Comparison of ankle-foot orthoses with plantar flexion stop and plantar flexion resistance in the gait of stroke patients: a randomized controlled trial. Prosthet Orthot Int. 2018;42(5):544–53.

    Article  Google Scholar 

  19. Waterval NFJ, Breham MA, Harlaar J, Nollet F. Description of orthotic properties and effect evaluation of ankle foot orthoses in non-spastic calf muscle weakness. J Rehabil Med. 2020;52:jrm00026.

    PubMed  Google Scholar 

  20. Kerkum YL, Houdijk H, Brehm MA, Buizer A, Kessels MLS, Sterk A, van den Noort JC, Harlaar J. The shank-to vertical angle as a parameter to evaluate tuing of ankle-foot orthoses. Gait Posture. 2014;42:269–74.

    Article  Google Scholar 

  21. De Seze MP, Bonhomme C, Daviet JC, Burguete B, Machat H, Rousseaux M. Effect of early compensation of distal motor deficiency by the Chignon ankle-foot orthosis on gait in hemiplegic patients: a randomized pilot study. Clin Rehabil. 2011;25(11):989–98.

    Article  Google Scholar 

  22. Fickey SN, Browne MG, Franz JR. Biomechanical effects of augmented ankle power output during human walking. J Exp Biol. 2018;221:Jeb1821113.

    Google Scholar 

  23. Bregman DJJ, De Groot V, Diggele PV, Meulman H, Houdijk H, Harlaar J. Polypropylene ankle foot orthoses to overcome drop-foot gait in central neurological patients: a mechanical and functional evaluation. Prosthet Orthot Int. 2010;34(3):293–304.

    Article  Google Scholar 

  24. Harlaar J, Brehm M, Becher JG, Bregmen J, Buurke J, Holtkamp F, DeGroot V, Nollet F. Studies examining the efficacy of ankle foot orthoses should report activity level and mechanical evidence. Prosthet Orthot Int. 2010;34:327–35.

    Article  Google Scholar 

  25. Neptune RR, Kautz SA, Zajac FE. Contributions of the individual ankle plantar flexors to support, forward progression and swing initiation during walking. J Biomech. 2001;34:1387–98.

    Article  CAS  Google Scholar 

  26. Kobayashi T, Orendurff MS, Hunt G, Lincoln LS, Gao F, LeCursi N, Foreman KB. An articulated ankle-foot orthosis with adjustable plantarflexion resistance, dorsiflexion resistance and alignment: a pilot study on mechanical properties and effects on stroke hemiparetic gait. Med Eng Phys. 2017;44:94–101.

    Article  Google Scholar 

  27. Nagaya M. Shoehorn-type ankle-foot orthoses: prediction of flexibility. Arch Phys Med Rehabili. 1997;78:82–4.

    Article  CAS  Google Scholar 

  28. Sumiya T, Suzuki Y, Kasahara S. Stiffness control in posterior-type plastic ankle-foot orthoses: effect of ankle trimline Part2: orthosis characteristics and orthosis/patient matching. Prosthet Orthot Int. 1996;20:132–7.

    Article  CAS  Google Scholar 

  29. Desloovere K, Molenaers G, Van Gestel L, Huenaerts C, Van Campenhout A, Callewaert B, Vande Walle P, Seyler J. How can push-off be preserved during use of an ankle foot orthosis in children with hemiplegia? A prospective controlled study. Gait Posture. 2006;24:142–51.

    Article  Google Scholar 

  30. Yamamoto S, Fuchi M, Yasui T. Change of rocker function in the gait of stroke patients using an ankle foot orthosis with an oil damper: immediate changes and the short-term effects. Prosthet Orthot Int. 2011;35(4):350–9.

    Article  Google Scholar 

  31. Polese JC, Teixeira-Salmela LF, Nascimento LR, Faria CDM, Kirkwood RN, Laurentino GC, Ada L. The effects of walking sticks on gait kinematics and kinetics with chronic stroke survivors. Clin Biomech. 2012;27:131–7.

    Article  Google Scholar 

Download references

Acknowledgements

The authors thank all patients who participated in this study and the physiotherapists who assisted with gait training and obtaining measurements.

Funding

This study was supported by JSPS KAKENHI grant number 17K01577.

Author information

Authors and Affiliations

Authors

Contributions

SY, NM, YK, and YO designed the study. NM and YK recruited patients and collected the data. SY, ST, and AD analyzed the data. SY drafted the manuscript, and AD revised the final version. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Sumiko Yamamoto.

Ethics declarations

Ethics approval and consent to participate

The study protocol was approved by the ethics committees of the International University of Health & Welfare and Nakaizu Rehabilitation Center. Written informed consent was obtained from all participants prior to their participation in this study.

Consent for publication

All authors give the permission of the publication of the article to the publisher.

Competing interests

The authors are unaware of any conflicts of interest concerning this study.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Yamamoto, S., Motojima, N., Kobayashi, Y. et al. Ankle–foot orthosis with an oil damper versus nonarticulated ankle–foot orthosis in the gait of patients with subacute stroke: a randomized controlled trial. J NeuroEngineering Rehabil 19, 50 (2022). https://doi.org/10.1186/s12984-022-01027-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12984-022-01027-1

Keywords