Gait training with a wearable powered robot during stroke rehabilitation: a randomized parallel-group trial

Abstract

Background

We have developed a wearable rehabilitation robot, “curara®,” and examined its immediate effect in patients with spinocerebellar degeneration and stroke, but its rehabilitative effect has not been clarified. The purpose of this study was to examine the effect of this device on gait training in stroke patients.

Methods

Forty stroke patients were enrolled in this study. The participants were divided randomly into two groups (groups A and B). The participants assigned to group A received RAGT with curara® type 4, whereas those in group B received conventional therapist-assisted gait training. The clinical trial period was 15 days. The participants performed 10 sessions of gait training (5 times per week) each lasting 30 ± 5 min per day. The 10-m walking time (10mWT), and 6-minute walking distance (6MWD) were evaluated as the main outcomes. Timed up and go and Berg Balance Scale (BBS) were also examined. Gait parameters (stride duration and length, standard deviation of stride duration and length, cadence, ratio of the stance/swing phases, minimum/maximum knee joint angle, and minimum/maximum hip joint angle) were measured using a RehaGait®. The items other than BBS were measured on days 0, 7, and 14, whereas BBS was measured on days 0 and 14. The improvement rate was calculated as the difference of values between days 14 and 0 divided by the value on day 0. The improvement rates of the 10mWT and 6MWD were set as the main outcomes.

Results

The data of 35 participants were analyzed. There was no significant difference in the main outcomes between both groups at the end of gait training. As for intragroup changes, gait speed, stride length, stride duration, and cadence were improved significantly between days 0 and 14 in each group. When examining the interaction effect between the day of measurement and group, stride duration (p = 0.006) and cadence (p = 0.012) were more significantly improved in group A than in group B.

Conclusions

This novel wearable powered robot may have the potential to improve gait speed of individuals in stroke rehabilitation.

Trial registration

Japan Registry of Clinical Trials (jRCTs032180163). Registered on February 22, 2019; https://jrct.niph.go.jp/en-latest-detail/jRCTs032180163.

UMIN Clinical Trials Registry (UMIN000034237)

Registered on September 22, 2018; https://center6.umin.ac.jp/cgi-open-bin/icdr/ctr_view.cgi?recptno=R000038939.

Background

Recently, robot-assisted gait training (RAGT) has been applied widely to individuals with stroke to regain and improve walking ability [1]. In the 1990s, body weight-supported treadmill training was introduced in the clinical setting [2]. Currently, rehabilitation robots with different assistive forms for lower limb movements have become popular, e.g., Gait Trainer® and Haptic Walker®, which assist with foot movements, and Lokomat®, which assists with lower limb orthosis on a treadmill. These devices enable the joints of the lower limbs to move in a state close to normal during walking without voluntary efforts from the patient [3,4,5].

In addition to stationary rehabilitation robots, wearable rehabilitation robots have been developed recently and used widely in gait training, e.g., Hybrid Assistive Limb (HAL®) and ReWALK® [6]. Generally, wearable rehabilitation robots are smaller and lighter than stationary rehabilitation robots; therefore, a great advantage of these devices is that individuals wearing them can move around freely. This makes it possible for people with a wearable rehabilitation robot to perform overground training in daily life. Systematic reviews have reported that rehabilitation robots improve balance and ankle spasticity in patients with brain diseases including stroke [7, 8]. Concerning the rehabilitative effect of HAL®, the degree of walking independence evaluated by the Functional Ambulation Category was significantly improved in the HAL®-wearing group compared to the non-wearing group [9, 10]. In addition, there are reports that RAGT showed more significant improvement in the time up and go (TUG) test, 10mWT and 6MWD than conventional walking training [11, 12].

We started clinical research in 2017 for the practical application of a wearable rehabilitation robot, “curara®”. We have examined the effect of the device in patients with spinocerebellar degeneration [13, 14] and stroke [15]. We have shown that the use of the device improves walking speed, stride length, walking rate, and asymmetry in stroke patients temporarily [15], but we have not yet clarified its rehabilitative effect on gait training. Therefore, the purpose of this study was to examine the effect of the device on gait training in stroke patients.

Methods

Participants and instrumentation

A total of 40 individuals participated in this study. They fulfilled all of the inclusion criteria and did not fulfill any of the exclusion criteria. The inclusion criteria were: (1) cerebrovascular disease presenting with hemiplegia; (2) aged ≥ 20 years; (3) 14–90 days after stroke onset; (4) able to walk ≥ 10 m independently with or without a walker and/or brace; and (5) Berg Balance Scale (BBS) score ≥ 26. The exclusion criteria were: (1) possible causes for gait disturbance other than cerebrovascular disease; (2) too thin or obese to fit into the device; and (3) any other reasons that were considered to make the subjects ineligible to participate (for example, severe dementia or psychiatric symptoms, severe spasticity or joint contracture of the paralyzed leg, etc.).

The participants were divided randomly into two groups (groups A and B). The participants in group A were assigned RAGT with the device, whereas those in group B received conventional therapist-assisted gait training. Treatment was randomly assigned at the Data Center of Center for Clinical Research in Shinshu University Hospital. The randomization list has been generated using Viedoc 4 ™ (Pharma Consulting Group, Sweden) that was used as an electronic data capture system for this study. To balance the two treatment arms, dynamic randomization performed between the two treatment arms in a ratio of 1:1 applying the modified algorithm of Pocock and Simon (as implemented in Viedoc). Characteristics of the participants were shown in Table 1. Baseline participants were 65.1 ± 12.9 years [mean ± standard deviation (SD)] in group A and 63.0 ± 12.9 in group B. Brunnstrom Recovery Stage scores were III - VI in Group A and IV - VI in group B. FIM and BBS scores were not different between groups A and B (FIM: 74.8 ± 12.6 vs. 75.6 ± 10.2; BBS: 52.1 ± 4.3 vs. 49.1 ± 6.8, respectively).

Table 1 Characteristics of the participants

We used a wearable “curara® type 4” robot in this study, which weighs approximately 5 kg (four actuator units and the control box) and has a novel structure that does not have direct rigid link between the actuator units on the hip and knee joints. Thus, both the actuator units can function independently. The basic mechanisms of the device are characterized by a torque-sensing technique and synchronized-based control system. Detailed information regarding the device is described in our previous reports [13, 15].

Outcome measures

The study period was 15 days in total. We evaluated the participants on days 0, 7, and 14. On these days, they were instructed to walk 10 m on a flat floor at a comfortable speed 9 times while wearing a RehaGait® analyzer (HASOMED, Magdeburg, Germany). We measured the 10-m walking time (10mWT) with a stopwatch and the distance walked in 6 min (6-minute walking distance; 6MWD). We also performed timed up and go (TUG) test on days 0, 7, and 14, and BBS on days 0 and 14. We collected the following gait parameters using RehaGait®: stride duration and length, standard deviation of stride duration and length, cadence, ratio of the stance/swing phases, minimum/maximum knee joint angle, and minimum/maximum hip joint angle. We set the improvement rates of the 10mWT and 6MWD as the main outcome measures, which were calculated as the difference of values between days 14 and 0 divided by the value on day 0. All of the measurements shown above were acquired without wearing the device.

Rehabilitation program

The participants performed 10 sessions of gait training (5 times per week) each lasting 30 ± 5 min session per day. One or two physical therapists accompanied each participant during gait training. For group A, the therapists operated the device and prevented falls, but they did not give any advice on gait to the participants. The participants also received a combination of physical (except gait training), occupational, or speech therapy rehabilitation. The maximum rehabilitation time including gait training was 3 h/day in both groups.

We set synchronization gain, gait cycle, and joint angles as the assist conditions of the device. Among these, synchronization gain was fixed to 0.1 at the hip joint and 0.3 at the knee joint throughout the rehabilitation period in all participants. The gait cycle and joint angles varied from individual to individual, and they were set according to the gait parameters of the fastest gait performance on days 0 and 7. The amplitude of the joint angle was set at 140% at the unaffected hip joint and 110% at the unaffected knee joint, and the gait cycle was set at 85%. Therefore, the assist conditions were set 40% wider hip joint, 10% wider at the knee joint, and 15% faster in the gait cycle than the gait parameters of the fastest gait performance. The conditions set on day 0 were used for gait training on days 1–6, and those updated on day 7 were utilized on days 8–13. These parameters meant that the amount of assistance provided by the robot was larger at the hip joint than at the knee joint, i.e., the device-in-charge robotic support was more influential at the hip joint than at the knee joint. With these assist conditions, the device enabled the participants to reproduce their best gait performance faithfully during gait training.

Statistical analysis

The distribution of characteristics of the participants between two groups was analyzed by a t-test for continuous variables and a chi-square test for categorical data. The differences of the BBS score and TUG test between days 0 and 14 were analyzed by a paired t-test. The differences of the gait parameters obtained with RehaGait® were analyzed using a generalized linear mixed model with Bonferroni’s correction. In the model, the day of measurement, group, and the interaction between those two factors were set as the fixed effects, and subject factors and the number of measurements on the same day of measurement were set as random effects. All statistical analyses were performed using IBM SPSS Statistics 24 for Windows. The level of significance was set at p < 0.05 in all tests.

Results

Of the 40 participants, three in group A and one in group B dropped out of the trial; two participants in group A retracted informed consent because of mental health issues and one discontinued the trial due to robot-induced skin problems, while one participant in group B had a second stroke attack during the trial period. A total of 36 participants (17 in group A and 19 in group B) completed the program without any adverse events. Detailed information on the participants is shown in Table 1. However, one participant in group B was excluded from the analysis due to extreme outliers in most of the measured items; therefore, the data of 35 participants were analyzed.

The results of the main outcome measures are shown in Table 2. The mean improvement rates of the 10mWT were 20.6% in group A and 13.9% in group B. However, the difference between both groups was not statistically significant (p = 0.099). The mean improvement rates of the 6MWD were 16.8% in group A and 19.4% in group B, again with no significant difference between both groups (p = 0.067). The amount of change in the BBS or TUG was not statistically significant between groups A and B.

Table 2 Results of the main outcomes, BBS, and TUG

The values of the gait parameters obtained by RehaGait® are shown in Table 3. Almost all of the gait parameters, except maximum flexion angles of the hip and knee joints, improved on day 14 in each group. When examining the interaction effect between the day of measurement and group, stride duration (p = 0.006) and cadence (p = 0.012) were more significantly. Therefore the results improved in group A than in group B. These results were expected to be reflected in gait speed, and the interaction effect of gait speed was very close to the level of statistical significance (p = 0.055).

Table 3 Gait parameters obtained with RehaGait®

The index of symmetry was expressed as the ratio of the maximum hip flexion angle of the affected leg to that of the unaffected leg in each group (Table 3). However, there was no significant difference in the index of symmetry within or between the groups.

Discussion

Seventeen out of 20 participants tolerated and coped well with a 15-day clinical trial with the device without any adverse events. The primary outcome measures in this study showed significant improvements (10–20% of the baseline value) in both groups; however, contrary to our expectations, there were no significant differences between the groups. The same outcome was found for the BBS and TUG.

When analyzing the RehaGait® data, significant intragroup differences in several parameters between days 0 and 14 were found in both groups. In particular, stride duration and cadence showed a more significant improvement in group A compared to the group B (interaction effect: p = 0.006 for stride duration, p = 0.012 for cadence). As a result, the interaction effect for gait speed almost reached statistical significance (p = 0.055). We reported previously that stride duration and cadence are increased in stroke patients when they walked with the device [15]. Thus, the rehabilitative effect of the device on stride duration and cadence was confirmed in the present study. This effect is very likely to contribute to the improvement of walking speed, and the assistance provided by the device, which shortened the gait cycle and enlarged the joint angles, would have led to this favorable change in stride duration and cadence.

In this study, all the participants were rather mild in the severity of hemiparesis at baseline because they had to train with the device (~ 5 kg) burdened when they were randomized to group A. In fact, the averaged gait speed was 0.90 m/sec at baseline (n = 36), which was faster than the gait speed (0.08–0.64 m/s) in the participants reported in the meta-analysis study investigating the RAGT effect [16]. In our preliminary study on stroke patients, which was a single-use analysis of curara® type 3, the average walking speed of the subjects was 0.4 m/s (n = 15) [15]. In this situation, we have to take the ceiling effect into consideration when interpreting the data, which is one of the major limitations of this study. As we measured the difference between the 10mWT on days 0 and 14, the ceiling effect should be minimized. The actual change in the 10mWT in group A (-0.23 ± 0.12 m/s, mean ± standard deviation) was greater than that in group B (-0.14 ± 0.12 m/s), but the difference between the groups was not statistically significant.

Another concern was whether or not the assistance provided by the device was best for each participant. We did not set synchronization gain at the hip and knee joints according to individual gait performance. Synchronization gain may be related to error learning that involves the trial and error required for motor learning [17]. A lower synchronization gain means a reduced degree of flexibility of intentional joint movement. Ideally the synchronization gain of the device should be set individually, but it is a complicated and time-consuming process to perform during the limited time available for this clinical trial.

During the study, we felt that the affinity of the participants to the robot varied from individual to individual. Roughly speaking, the more elderly individuals tended to have less affinity for the device than the younger ones, and individuals with better walking ability had lower motivation to practice with the robot. One factor influencing affinity to the robot was that RAGT is still being developed as a method for stroke rehabilitation throughout Japan. For RAGT to become more popular, high-quality evidence should be acquired in future randomized case-control studies.

At least, we can say that RAGT with curara® had a rehabilitative effect identical to that of therapist-assisted gait training. Our initial motive was to develop a wearable robot that elderly or handicapped people can use for gait rehabilitation in daily life. The results shown above are encouraging for us to improve the devise further. First, we need to improve its operability and user friendliness so that anybody can use it easily. At the present time, we are creating an advanced type of robot using the experience and knowledge acquired in this study. Furthermore, we need to identify the best application timing and assist conditions for the device in stroke patients.

Conclusion

RAGT with our novel wearable powered robot may help to improve the walking speed of stroke patients. In order to use the device for walking rehabilitation in daily life, we have to review the selection criteria and outcomes of the subjects in future studies and further improve the operability and usability of the device.

Data Availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

6MWD:

6-minute walking distance

10mWT:

10-m walking time

BBS:

Berg Balance Scale

RAGT:

robot-assisted gait training

TUG:

timed up and go

References

  1. Ha YK, Joon HS, Sung PY, Min AS, Stephanie HL. Robot-assisted gait training for balance and lower extremity function in patients with infratentorial stroke: a single-blinded randomized controlled trial. J Neuroeng Rehabil. 2019;16:99.

    Article  Google Scholar 

  2. Finch IL, Barbeau H, Arsenault B. Influence of body weight support on normal human gait: development of a gait retraining strategy. Phys Ther. 1991;71:842–55.

    Article  CAS  PubMed  Google Scholar 

  3. Hesse S, Uhlenbrock D. A mechanized gait trainer for restoration of gait. J Rehabil Res Dev. 2000;37:701–8.

    CAS  PubMed  Google Scholar 

  4. Andreas M, Markus K, Ellen Q, Heinz M, Katrin F, Leopold S. Prospective, blinded, randomized crossover study of gait rehabilitation in stroke patients using the Lokomat gait orthosis. Neurorehabil Neural Repair. 2007;21:307–14.

    Article  Google Scholar 

  5. Schmidt H, Hesse S, Bernhardt R, Krüger J. Haptic Walker a novel haptic foot device.April 2005ACM Transactions on Applied Perception2:166–180.

  6. Uriel MH, Benjamin M, Tareq A, Leen J, James M, Dingguo Z. Wearable Assistive Robotics: a perspective on current Challenges and Future Trends. Sens (Basel). 2021;21:6751.

    Article  Google Scholar 

  7. Qing XZ, Li G, Carol CW, Qi SM, Yan TL, Ping PH, et al. Robot-assisted therapy for balance function rehabilitation after stroke: a systematic review and meta-analysis. Int J Nurs Stud. 2019;95:7–18.

    Article  Google Scholar 

  8. Shakti D, Mathew L, Kumar N, Kataria C. Effectiveness of robo-assisted lower limb rehabilitation for spastic patients: a systematic review. Biosens Bioelectron. 2018;117:403–15.

    Article  CAS  PubMed  Google Scholar 

  9. Watanabe H, Tanaka N, Inuta T, Saitou H, Yanagi H. Locomotion improvement using a hybrid assistive limb in recovery phase stroke patients: a randomized controlled pilot study. Arch Phys Med Rehabil. 2014;95:2006–12.

    Article  PubMed  Google Scholar 

  10. Watanabe H, Goto R, Tanaka N, Matsumura A, Yanagi H. Effects of gait training using the Hybrid Assistive Limb® in recovery-phase stroke patients: a 2-month follow-up, randomized, controlled study. NeuroRehabilitation. 2017;40:363–7.

    Article  PubMed  Google Scholar 

  11. Michela G, Chiara I, Emanuele R, Sanaz P, Chiara DB, Daniele G, et al. Stroke gait rehabilitation: a comparison of end-effector, overground exoskeleton, and conventional gait training. Appl nces. 2019;9:2627.

    Google Scholar 

  12. Franco M, Eleonora G, Michela G, Rocco SC, Sanaz P, Marina G, et al. Gait recovery with an overground powered exoskeleton: a randomized controlled trial on subacute stroke subjects. Brain Sci. 2021;11:104.

    Article  Google Scholar 

  13. Tsukahara A, Yoshida K, Matsushima A, Ajima K, Kuroda C, Mizukami N, et al. Effects of gait support in patients with spinocerebellar degeneration by a wearable robot based on synchronization control. J NeuroEng Rehabili. 2018;15:84.

    Article  Google Scholar 

  14. Matsushima A, Maruyama Y, Mizukami N, Tetsuya M, Hashimoto M, Yoshida K. Gait training with a wearable curara® robot for cerebellar ataxia: a single-arm study. Biomed Eng Online. 2021;20:90.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Mizukami N, Takeuchi S, Tetsuya M, Tsukahara A, Hashimoto M, Yoshida K, et al. Effect of the synchronization-based control of a wearable robot having a non-exoskeletal structure on the hemiplegic gait of stroke patients. IEEE Trans Neural Syst Rehabil Eng. 2018;26:1011–6.

    Article  PubMed  Google Scholar 

  16. Geoffroy M, Romain G, David G, Bertrand G, Laurent B, Patrick D, et al. Effects of robotic gait training after stroke: a meta-analysis. Ann Phys Rehabil Med. 2020;63:518–34.

    Article  Google Scholar 

  17. Joseph H, Diane N, Marlena P, Kathy B, Donielle DC, Jennifer HK, et al. Multicenter randomized clinical trial evaluating the effectiveness of the Lokomat in subacute stroke. Neurorehabil Neural Repair. 2009;23:5–13.

    Article  Google Scholar 

Download references

Acknowledgements

We thank to all the participants in this study and physical therapists and occupational therapists (JA Nagano Koseiren Kakeyu-Misayama Rehabilitation Center Kakeyu Hospital) for their professional help in RAGT and therapist-assisted gait training. We are also grateful to Ms. Yoshiko Takagi, Mr. Satoshi Hokari, Mr. Takashi Igarashi (The Center for Clinical Research, Shinshu University Hospital) for data management and monitoring of the study. A draft of this manuscript was edited by NAI, Inc. (Yokohama, Japan)

Funding

This study was supported by the Japan Agency for Medical Research and Development (AMED) under Grant Number JP17hk0102048-19hk0102048.

Author information

Authors and Affiliations

Authors

Contributions

AM, MH and KY conceived the study. AM, YM, NM, MT, MH and KY contributed to study design. DM, AM, YM and KY contributed to data collection. DM and AM performed statistical analysis. DM, AM, YM and KY interpreted data. DM, AM and KY drafted the manuscript. DM, AM, YM, NM, MT, MH and KY revised and approved the final manuscript.

Corresponding author

Correspondence to Kunihiro Yoshida.

Ethics declarations

Ethics approval and consent to participate

The study procedure was approved by the Ethics Committee of Shinshu University School of Medicine (No. 3999), and registered in the Japan Registry of Clinical Trials (Trial ID 032180163: https://jrct.niph.go.jp/en-latest-detail/jRCTs032180163) on February 27, 2019, and in the UMIN Clinical Trials Registry (UMIN000034237: https://center6.umin.ac.jp/cgi-open-bin/icdr/ctr_view.cgi?recptno=R000038939) on September 22, 2018. All participants were provided with all necessary information about the study and provided written informed consent before the clinical trial. All aspects of the study conformed to the principles described in the Declaration of Helsinki.

Consent for publication

Not applicable.

Competing interests

Mr, Miyagawa, Dr. Matsushima, Mr. Maruyama and Mr. Tetsuya have no competing interests.

Dr. Mizukami is funded by the Japan Society for the Promotion of Science (Grant-in-Aid for Young Scientists).

Dr. Mizukami is funded by the Japan Society for the Promotion of Science grants (Grant-in-Aid for Scientific Research (C) and Challenging Research [Exploratory]), and was funded by the Japan Society for Promotion of Science grants (Grant-in-Aid for Scientific Research [B]), Tokyo Metropolitan Industrial Technology Research Institute grants (Robot industry activation business), and AMED grants (Robot care equipment development and standardization business).

Dr. Yoshida is funded by Health and Labour Sciences Research Grants (Research Committee of the Ataxia, Research on Policy Planning and Evaluation for Rare and Intractable Diseases, The Ministry of Health, Labour, and Welfare, Japan).

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Miyagawa, D., Matsushima, A., Maruyama, Y. et al. Gait training with a wearable powered robot during stroke rehabilitation: a randomized parallel-group trial. J NeuroEngineering Rehabil 20, 54 (2023). https://doi.org/10.1186/s12984-023-01168-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12984-023-01168-x

Keywords