Characterizing the supraspinal sensorimotor control of walking using MRI-compatible system: a systematic review

Abstract

Background

The regulation of gait is critical to many activities of everyday life. When walking, somatosensory information obtained from mechanoreceptors throughout body is delivered to numerous supraspinal networks and used to execute the appropriate motion to meet ever-changing environmental and task demands. Aging and age-related conditions oftentimes alter the supraspinal sensorimotor control of walking, including the responsiveness of the cortical brain regions to the sensorimotor inputs obtained from the peripheral nervous system, resulting in diminished mobility in the older adult population. It is thus important to explicitly characterize such supraspinal sensorimotor elements of walking, providing knowledge informing novel rehabilitative targets. The past efforts majorly relied upon mental imagery or virtual reality to study the supraspinal control of walking. Recent efforts have been made to develop magnetic resonance imaging (MRI)-compatible devices simulating specific somatosensory and/or motor aspects of walking. However, there exists large variance in the design and functionality of these devices, and as such inconsistent functional MRI (fMRI) observations.

Methods

We have therefore completed a systematic review to summarize current achievements in the development of these MRI-compatible devices and synthesize available imaging results emanating from studies that have utilized these devices.

Results

The device design, study protocol and neuroimaging observations of 26 studies using 13 types of devices were extracted. Three of these devices can provide somatosensory stimuli, eight motor stimuli, and two both types of stimuli. Our review demonstrated that using these devices, fMRI data of brain activation can be successfully obtained when participants remain motionless and experience sensorimotor stimulation during fMRI acquisition. The activation in multiple cortical (e.g., primary sensorimotor cortex) and subcortical (e.g., cerebellum) regions has been each linked to these types of walking-related sensorimotor stimuli.

Conclusion

The observations of these publications suggest the promise of implementing these devices to characterize the supraspinal sensorimotor control of walking. Still, the evidence level of these neuroimaging observations was still low due to small sample size and varied study protocols, which thus needs to be confirmed via studies with more rigorous design.

Background

The regulation of gait is critical to many activities of everyday life. Aging and age-related conditions (e.g., movement disorders [1], stroke [2], chronic pain [3], etc.) oftentimes diminish gait performance [4, 5] and thus increase the risk of deconditioning, mobility decline, and falls [6, 7] in the older adult population and those suffering from gait disorders (e.g., shuffling gait). Gait is regulated by a complex system that requires ongoing communication between peripheral neuromuscular circuitry [8] and numerous subcortical and cortical networks [9]. In particular, when walking, somatosensory information obtained from mechanoreceptors throughout the body is delivered via peripheral nerves to the central nervous system, where it is processed and integrated with visual and vestibular feedback in numerous cortical and sub-cortical networks and used to develop and execute appropriate motor programs to meet ever-changing environmental and task demands [10]. In addition to the issues in peripheral musculoskeletal systems (e.g., osteoporosis), age and age-related conditions (e.g., stroke and Parkinson’s disease) oftentimes alters the functional characteristics of supraspinal elements (e.g., the brain cortical regions) of walking, resulting in gait disorders. Considerable effort has therefore been placed on the study of the supraspinal control of walking in order to understand the impact of aging and age-related conditions on locomotor control and to identify new targets for preventive and rehabilitative medicine in those suffering from diminished mobility.

Functional magnetic resonance imaging (fMRI) enables the characterization of neural activity with high spatial resolution and can thus provide insight into the supraspinal control of many important functions and behaviors [11]. However, it is challenging to utilize MRI to characterize locomotor control because the individual’s head is required to stay motionless throughout the scan. The majority of efforts to date have thus relied upon mental imagery to study the planning and thought of carrying out movements [12,13,14], or virtual reality to study the processing of visual feedback related to navigating an environment [15,16,17,18]. This research, while important, provides very little information regarding brain-level, somatosensory processing or the sensorimotor control involved in walking [14, 19]. As a result, recent efforts have been made to develop MRI-compatible devices that simulate specific somatosensory and/or motor aspects of walking, while limiting movement of the head. Studies using these devices have indeed provided new insights to locomotor control. At the same time, however, there exists large variance in the design and functionality of these devices, and as such, inconsistent fMRI observations. We have therefore completed a systematic review to (1) summarize the design and functionality of MRI-compatible devices that have been developed to characterize the supraspinal processing of sensorimotor aspects of locomotion, and (2) synthesize available BOLD results emanating from studies that have utilized these devices. The knowledge provided by this work is expected to inform the design of future studies implementing these devices to characterize the supraspinal elements of gait regulation in those suffering from gait disorders.

Method

Search strategy

A systematic literature search of PubMed, Web of science, EBSCO MEDLINE, SPORT Discus, Psych-Info, Cochrane library and Scopus was performed with the last search completed on August 27, 2022. The search field focused on the magnetic resonance imaging compatibility of the devices (i.e., MRI-compatible) and the synonyms for gait including lower-extremity motor paradigms that were considered as the clinical surrogates of gait. The following search terms were used to identify relevant literature in the database: (“magnetic resonance imaging compatible” or “MRI compatible”) and (“gait” OR “walk” OR “step*” OR “ambul*” OR “locomot*” OR “lower limb movement” OR “pedal” OR “dorsiflexion” OR “plantarflexion” OR “ankle motion*” OR “foot sole pressure stimul*”). A manual search of the bibliographic references of extracted articles and existing reviews was also conducted to identify studies that were not captured in the electronic searches.

Inclusion and exclusion criteria

The inclusion criteria were: (1) the device used in the study was MRI-compatible with mechanical structure enabling the control or adjustment of stimulation parameters (e.g., applied force, movement speed of stimulation); (2) the goal of the study or device development was to explore the functional characteristics of the supraspinal regions pertaining to walking-related sensorimotor stimuli. Manuscripts were excluded if they: (1) only contained a descriptive overview of the device design; (2) were systematic reviews, case reports, protocol papers, conference abstracts, or letters to the editor; and (3) were not written in English.

Data extraction and synthesis

Two authors (YH and JZ) independently performed data extraction, and when disagreement on the extraction of data/information was present, it was discussed with additional authors (DB and BM) until a consensus was achieved. The following data were extracted for each study: (1) device characteristics: simulation protocol of sensorimotor characteristics of gait, design of power supply, execution unit and control unit, materials used to develop the devices, accessories to minimize unwanted or extraneous motions, and the results of MRI compatibility test and the validation test of device performance; and (2) and study characteristics: the information of the study team (e.g., first author, year of publishing, country), participant characteristics, fMRI protocol and design, and the results of head-motion artifacts and fMRI BOLD signal analysis.

Results

Study selection

The data summary and analysis were completed on November 10th, 2022. Figure 1 presents the flow diagram of study selection. Our initial search retrieved 1,027 articles from four databases and 13 additional articles from other sources (e.g., reference lists from original work and review articles). After the removal of duplicates and the screening of title, abstract, and full text for study design and outcomes, 26 original research articles were eligible and included in the systematic review. Reasons for exclusion in this phase included: not related to gait, not an fMRI study, prototype description only, or not an MRI-compatible device with mechanical structure enabling the control or adjustment of stimulation parameters.

Fig. 1
figure 1

The RISMA flowchart of publication screening. Abbreviation PRISMA: Preferred Reporting Items for Systematic Reviews and Meta-Analyses; fMRI: functional magnetic resonance imaging

Device feasibility

A total of 13 different devices were reported in 26 total studies, including a foot-sole stimulation system [20], a dual-drive foot-sole stimulator [21], a Korvit boot system [22,23,24], a foot pedal manipulandum [25], a plantar flexion force measure apparatus [26], a brain discovery pneumatic orthosis (Bra.Di.P.O.) [27, 28], a bipedal device (i.e., one pedal for each of left and right foot) [29], a torque-measuring apparatus [30], a pedaling device [31,32,33], a pseudogait-magnetic resonance compatible device (pseudogait-MRCD) [34, 35], a cylindrical treadmill device [36], a magnetic resonance compatible stepper (MARCOS) [37,38,39,40,41], and a lower-extremity motion simulator (LOMS) [42,43,44,45] (Table 1).

Table 1 Device design and functionality

Device design

The structure of included devices can be divided into three main parts: the power supply unit, the execution unit that is oftentimes placed in the MRI scan room, and the control unit placed outside the scan room (Fig. 2; Table 2).

Fig. 2
figure 2

The conceptual diagram of device structure

Table 2 Information on the components of device

For power supply, four devices used an air compressor to drive a pneumatic actuator [20,21,22,23,24, 42,43,44,45], while such information is missing for the other nine.

The execution units consist of three structures: support to the lower limb, sensor, and actuator. Types of lower limb support included a plastic boot (number of devices (N) = 3) [20,21,22,23,24], syringes (N = 1) [26], a customized pedal (N = 5) [25, 27,28,29,30,31,32,33], a special treadmill (N = 2) [34,35,36], and a customized lower limb exoskeleton (N = 2) [37,38,39,40,41,42,43,44,45]. Sensor types included force/torque sensors (N = 2) [26, 30], displacement sensors (optical encoders [31,32,33,34,35] or a rotary potentiometer [42,43,44,45])(N = 3), and the combination of at least two types of sensors (e.g., the Bra.Di.P.O. used both a custom-built analogue optical encoder and a pressure transducer [27, 28])(N = 3) [27,28,29, 37,38,39,40]. Five devices did not report sensor information. For the types of actuators, pneumatic actuators were used in seven devices [20,21,22,23, 27, 28, 36,37,38,39,40,41,42,43,44,45], while such information was missing for the other six.

Control units consisted of valves, a controller, a data acquisition system, and a user interface. The details of valves were provided in five devices [20, 21, 27, 28, 37, 38, 42,43,44], and that of controller in four [20, 21, 37, 42, 44], data acquisition system in five [29,30,31,32, 34, 37], and user interface in six devices [20, 21, 30,31,32, 34, 37] (See Table 2 for details).

In order to prevent unwanted task-related movement of the body as well as the extraneous motion of head that may interfere with image quality, all devices implemented different accessories, such as Velcro straps or belts (N = 6) [26, 29, 31,32,33,34,35,36,37,38,39, 41], triangular platforms (N = 4) [26, 30, 34, 35], cushion (N = 1) [36], and/or support in the head coil of MRI scanner to help secure the head (i.e., pillows (N = 4) [26, 31,32,33,34,35, 37,38,39,40,41] or foam blocks (N = 3) [23,24,25, 31,32,33].

Device compatibility with fMRI

MRI compatibility of seven out of the thirteen devices were tested via phantom-based imaging quality tests and reported in nine publications [20, 21, 31, 34, 36, 37, 42, 43, 45]. Uniquely, the compatibility of one device was also tested when human participants were performing tasks (e.g., voluntary motion of the lower extremities with and without wearing the lower-extremity motion simulator ) [42, 43]. Specifically, two studies measured spatial signal-to-noise ratio (SNR) [31, 45]; two measured temporal SNR [36, 37]; one measured both spatial SNR and temporal SNR [34]; one measured spatial SNR, the signal-to-fluctuation-noise ratio (SFNR) and the field map mean [20]; one measured spatial SNR, SFNR and the field non-uniformities [21]; and the other two measured the signal intensity and dispersion [42, 43] (Table 3). Additionally, both non-metallic materials including plastic, rubber, and polyvinyl chloride (PVC) and non-magnetic metal including aluminum, bronze, brass, and stainless steel were used to manufacture devices to ensure the MRI compatibility (Table 3).

Table 3 Information on the compatibility of device

Device functionality

The overarching goal of included devices was to simulate walking-related somatosensory and/or motor experiences of walking over ground, during the MRI scan, thereby enabling study of related brain activity. As such, devices were designed to apply somatosensory stimuli to the foot soles (N = 3), induce locomotor-like movements of the lower extremities (including single-joint movement of ankle (N = 4), and multi-joint movements (N = 4)), or induce locomotor-like movements of the lower extremities while simultaneously applying somatosensory stimuli to the foot soles (N = 2). Additionally, three devices were designed for passive participant engagement, seven for active engagement to initiate movements by participants, and the other three for both functionalities (Table 1).

Three devices were designed to apply somatosensory stimuli to the soles of one or both feet [20,21,22,23,24] that mimic the ground reaction forces (GRF) experienced when walking. Specifically, the foot-sole stimulation system was designed to apply a sinusoidal pressure waveform with a peak force of 200 N and in a range of frequency from 1 to 10 Hz to one region of one foot sole [20]. Two other similar devices were designed to apply pressure stimuli to multiple areas of one or both soles: the dual-drive foot-sole stimulator was designed to apply programmable waveform-type stimuli reproducing the pressure waveforms of each participant as recorded during walking over an instrumented pressure insole prior to the scan [21]; the Korvit boot system was designed to apply two different forms of stimulation including “gait-like” and “chaotic” sequences [22,23,24].

Eight devices were designed to enable locomotor-like movements of the lower extremities. Five of these devices were designed to induce single-joint movement (i.e., ankle dorsiflexion and plantarflexion) and three were designed to induce multi-joint movements. Of the five single-joint devices, two were designed to enable unilateral movement [25, 30]. The other three were created to enable either unilateral or bilateral ankle dorsiflexion and plantarflexion with unique functions: one device provided real-time feedback of plantar flexion exertion in order to have participants attempt to achieve target forces [26]; one device was created to enable synchronization of the angle and torque characteristics of ankle movements in order to ensure consistency of task performance across trials [29]; and a final device (i.e., Bra.Di.P.O) was created to induce passive ankle motion patterns [27, 28]. Of the three devices enabling multi-joint movements, the pedal device was created to induce movements at different speeds without the restriction of knee motion [31,32,33]; the cylindrical treadmill device was created to enable participants to perform stepping movements by moving their legs to rotate a cylinder [36]; and the pseudogait-MRCD consisted of a vertical treadmill connected to a triangular platform that was create to enable participants to perform voluntary step-like movements [34, 35].

The final two devices were designed to enable locomotor-like movements of the lower extremities combined with extra somatosensory stimuli [26,27,28,29,30,31,32,33,34]. Specifically, the MARCOS device was described as being actuated by two pneumatic cylinders: one attached to knee orthoses allowing predefined flexion and extension movements of each leg in the sagittal plane, and the other attached to the shoe rendering external loads of up to 400 N along the cranio-caudal body axis to the foot-sole [37,38,39,40,41]. The LOMS device was created to enable three degree-of-freedom motion of each leg and simultaneously applies the ground reaction force of walking to foot soles via the foot stimulating parts [42,43,44,45].

Validation of device performance

Device performance was assessed in eight devices (Table 3). The similarity between the pressure as experienced during over ground walking and those simulated by the devices was tested for one device [21]; theoretical outputs (e.g., the output calculated from the formula using preprogrammed parameters) and the results obtained from the actual tests were compared in two devices [20, 28]; the performance of iterative learning in the controller for one (i.e., the shift/difference between the maximum phase of the expected and measured trajectories of position or force) [38]; and the reliability of motion (e.g., the mean amplitude of the torques) across different time points [29, 30, 40] and/or load levels [41] was tested for three. No such information was provided for the other five devices [22,23,24,25,26,27, 31,32,33,34,35,36, 39, 42,43,44,45].

The fMRI studies

Study characteristics

The 26 included fMRI studies were completed by 12 groups from nine countries. A total of 371 participants were included across studies. The sample size in the included studies ranged from one to 67. The sex of participants was reported in 17 of 26 studies. Twenty-five studies focused on healthy participants and one study focused on participants with brain damage due to stroke [33]. Ten studies reported both the age range and average age (one of which was median rather than mean [30]) of participants, five reported only the age range, seven reported only the average age (one of which had only one participant [37]), and four did not report age information (Table 4).

Table 4 The design of fMRI task and key findings

Only two of the 26 studies reported on safety testing. One reported that the safety of their device was examined in a previous iterative pilot testing, but did not report details of the safety test [29], and the other reported that it was confirmed that all components of the device were not attracted by a U-shaped neodymium magnet [45].

Fifteen of 26 studies reported information related to head-motion artifacts associated with usage of the devices [22, 25, 27, 28, 30, 39, 41,42,43,44,45]. Twelve of these studies reported that head motion was less than 3 mm [20, 21, 26, 29, 31,32,33,34,35,36,37,38]; the other three did not report the details of head motion, but stated that the data with excessive head motion were excluded from analyses [23, 24, 40] (Table 3).

Twenty-three of 26 studies utilized a block-design fMRI paradigm and the remaining three studies utilized event-related designs [25, 26, 37]. Eighteen of 26 studies reported the details of imaging processing procedure [20,21,22,23,24,25,26, 28,29,30,31,32,33,34,35,36, 39, 40], while such information is missing for the other eight. Specially, four of the 18 studies reported the utilization of the sparse sampling imaging protocol [31,32,33, 39].

For the comparisons performed in the included studies, 17 of 26 studies compared the brain activation between different stimulation conditions (e.g., unipedal versus bipedal stimulation, active versus passive movement) [22, 24,25,26,27,28,29,30, 32, 35,36,37,38, 40, 41, 43, 44], one between the condition of using mechanical stimulation induced by the device only and the motor imagery task only [23], one between stroke cohort and age-matched healthy counterparts [33] under the same stimulation condition, one between movements with and without the device stimulation [42], and the other six perform the comparison between task and blank blocks [20, 21, 31, 34, 39, 45].

The fMRI observations

The 26 included fMRI studies characterized the supraspinal activation in response to somatosensory stimulation, to locomotor-like movements of the lower extremities, or to locomotor-like movements of the lower extremities in combination with extra simultaneous somatosensory stimulation. Key fMRI results are presented in Table 4.

The activation of supraspinal regions in response to somatosensory stimulation, as compared to rest, were assessed in five studies [20, 22,23,24, 46]. Activation was observed in cortical regions, including primary sensorimotor cortex (S1/M1) (number of studies (n) = 4) [20,21,22,23], the secondary sensory cortex (S2) (n = 2) [20, 23], the supplementary motor area (SMA) (n = 4) [21,22,23,24], the pre-motor cortex (PMC) (n = 1) [22], the prefrontal cortex (PFC) (n = 2) [22, 24], and insula (n = 2) [21, 22], as well as in subcortical regions including the basal ganglia (n = 2) [22, 24], the thalamus (n = 1) [24] and the cerebellum (Cb) (n = 1) [22]. Specifically, Hao et al., observed that the single-point sinusoidal stimulation applied to foot soles induced the activation in S1,S2 and M1 [20]. By using a newer version of the stimulation system (i.e., dual-drive foot-sole stimulator), the same group confirmed such results and further demonstrated that the insula and cingulate cortex may play important role in the processing of gait-related somatosensory stimuli [21]. In the three studies using the Korvit boot system, Kremneva et al., observed the involvement of PFC in response to standing-related stimulation and a broad involvement of the primary and secondary sensorimotor cortices in response to slow-walking stimulation [22]. In another study, Labriffe et al., observed that both gait-like and non-gait-like stimulation can induce similar-amplitude activation of the primary sensorimotor cortex and secondary somatosensory cortex bilaterally [23]. Using the same device, Jeanvoine et al., further observed that older age was associated with greater activation of right pre-SMA and mid-dorsolateral PFC [24].

The activation of supraspinal regions in response to locomotor-like movements of the lower extremities were assessed in 21 studies. In seventeen of these studies, single-joint movement of ankle (n = 6) [25,26,27,28,29,30] or multiple-joint movements (n = 11) [31,32,33,34,35,36, 38, 39, 42, 44, 45] were implemented, and in the other four, the locomotor-like movements were applied simultaneously with somatosensory stimuli as controlled by the device (i.e., combined-type stimulation) [37, 40, 41, 43]. Across these studies, activation was observed within the S1/M1 and S2 (n = 4) [29, 30, 39, 40], SMA (n = 15) [25, 27, 28, 30,31,32,33,34, 37, 39, 40, 42,43,44,45], PMC (n = 6) [29,30,31, 33, 34, 38], insula (n = 1) [26], basal ganglia (n = 6) [25, 29, 30, 34, 35, 39], thalamus (n = 2) [25, 30] and/or Cb (n = 15) [25, 27,28,29,30,31,32,33,34,35,36,37, 43,44,45] (Table 4).

With respect to single-joint movements, Trinastic et al. [25], and Newton et al. [30], compared the brain activation in response to active (i.e., the participant performed the movement voluntarily) ankle dorsiflexion and plantarflexion. Both studies observed that the S1/M1 and SMA were activated by both movements, and uniquely, other brain regions (e.g., right putamen) were activated only by dorsiflexion. Doolittle et al. [29], compared the activation in response to active unilateral and bilateral motions of the ankle joint, and observed that the activation was not significantly different between these two conditions. In another study [26], the participant was asked to complete one trial each of plantarflexion of left foot, right foot, or both feet against 15% of his/her maximum voluntary contraction lasting for 5 s. It was observed that compared to unilateral (i.e., left or right foot) plantarflexion, bilateral plantarflexion induced greater activation in multiple cortical and subcortical regions (e.g., S1/M1 and PMC). Two studies directly compared the brain activation in response to active and “passive” (i.e., the device imposed participant movement) ankle movement. The results of both studies suggested that both active and passive movement induced activation in sensorimotor and supplementary motor regions, and active movements uniquely induced activation in thalamic, frontal and cingulated regions, while passive movements induced activation in temporal and parietal areas [27, 28].

With respect to multiple-joint movements, Promjunyakul et al. [33], completed a study consisting of people recovering from stroke. They observed that as compared to age-matched healthy controls, the activation induced by actively pedaling, especially in the SMA, PMA and Cb, was significantly lower in people recovering from stroke. Brain activation in response to active and passive multiple-joint movements was also examined in four studies, showing different observations. Jaeger et al. [39], observed that compared to passive movements, active movements induced higher activation of S1/M1, SMA-proper, cingulate motor area, S2, Cb and putamen; in another study, Mehta et al., observed that compared to passive movements, active movements induced higher activation within Cb only [32]; Hollnagel et al., observed [38] that active movements elicited activation in a more extensive sensorimotor network; and Takahiro et al. [44], observed that the sensory regions of the brain as activated by passive movement was broader than the areas activated by active movement. Additionally, the association between the walking-related supraspinal activation and the parameters of the movement (i.e., pace or frequency of the motion) was examined in three studies. Specifically, Mehta et al. [32], reported that higher pedaling rate was associated with greater activation in S1/M1, SMA, and Cb; Toyomura et al. [36], showed that slower speed of stepping elicited more-extensive activity in sensorimotor cortex and Cb; and Martinez et al. [35], observed that the increase of stepping frequency was associated with a decrease in the activation of Cb.

With respect to the combined-type stimulation (i.e., locomotor-like movements applied simultaneously with somatosensory stimuli as controlled by the device), Ikeda et al. [43], observed higher level of activation within the motor regions and Cb as induced by the locomotor-like movements alone, as compared to the combined-type stimuli. Likewise, Hollnagel et al. [37], reported observations in one participant that the combined-type stimuli induced lower activation in SMA and PMC as compared to locomotor-like movements alone. In another study, Jaeger et al. [41], examined supraspinal activation when participants were performing stepping at different levels of force simulating the GRF experienced during real walking (i.e., 0% (no force), 20% and 40% of the individual’s body weight). No significant difference of the activation in sensorimotor regions between different levels of force was observed. In the same study, the influences of active and passive movements were also compared. It was observed that across all levels of force applied to the foot soles, active movements induced higher activation in Cb compared to passive movements. In another study from the same team, it was observed that both active and passive stepping movements with a load of 40% of individual’s body weight induce activation in the bilateral S1/M1, SMA-proper and Cb [40].

Discussion

Thirteen different devices have been developed and implemented in fMRI studies. These previous works suggest a great promise to utilize these devices within the MRI environment to apply controlled walking-related stimuli to induce activation within multiple cortical and subcortical regions. Replication of results in many of the studies, however, was limited by small sample sizes, heterogeneity of the participant characteristics, and limited reporting of device validation, design, and fMRI compatibility. Future studies with larger sample size and rigorous design and validation testing are thus warranted to confirm and extend the observations from these publications on the characteristics of supraspinal control of gait in aging process and under the influences of age-related conditions via the implementation of these MRI-compatible devices.

In the included publications, device validation efforts have been made for stimulation performance of the devices, which are critical to the quality of neuroimaging data and the observations of brain activation. Several teams implemented adaptive control protocols based upon actuator performance [28], and/or utilized a iterative learning controller [38], to overcome the potential delay of work commands caused by the distance between the control unit (placed outside the scan room) and execution circuit [28, 48]. Future work (on pneumatic devices specifically) should consider working to reduce air loss in transmission via better sealing of the pistons, which would help to increase power and energy efficiency [49]. Additionally, between-subject variance in the characteristics of the lower extremity (e.g., different strength and lengths of the lower extremity between participants) may influence the device performance across individuals; and in turn, the degree to which a given device accurately simulates one or more elements of walking is a critical factor influencing fMRI results [21, 24]. It is thus recommended that future efforts to develop devices that can provide person-specific stimulation to more closely mimic their unique sensory or sensorimotor experience of walking, with the help of independently-controlled pneumatic actuators combined with pressure sensors to provide real-time feedback of applied pressure (i.e., closed-loop control of the stimulation) [21].

Efforts have also been made for the fMRI compatibility of these devices. For example, the execution units, which were placed inside the scan room, were uniquely manufactured with nonmagnetic and nonconductive materials with the goal to ensure sufficient robustness and to produce accurate and reliable stimulation to the individual during the MRI. Another challenge of using functional MRI to study brain activation induced by lower extremity stimulation is that stimulating or moving the legs and/or feet is likely to induce head motion artifact, which may in turn diminish image quality. Included studies successfully restricted movements of the head and trunk by using foam blocks and straps, and/or placing the knees in a flexed position and supporting the leg with a platform or cushion [26, 30, 34, 35]. Advanced signal processing techniques were also implemented to mitigate the influence of motion artifact on data quality in some studies. One commonly-used technique was “sparse sampling imaging” [50], which has been proven to effectively minimize movement-related artifacts on image quality [31,32,33, 39]. With the help of these strategies, studies that have reported the result of head-motion artifacts showed that the artifacts were less than 3 mm (in some cases even < 1 mm), which would not alter the observation of the brain activation [51]. It should be noted that most included studies only recruited healthy younger adult volunteer without overt neurological disorders (only one included individuals recovering from stroke), and the device performance and fMRI compatibility were not explicitly reported in seven of the included devices. It is thus worthwhile to explore and establish a thorough and standardized procedure to examine the device performance and MRI compatibility, and a validated protocol of data acquisition and processing, enabling the production of reproducible sensorimotor stimulation, and thus reliable observation of the brain’s activation across individuals [38, 48].

In general, across all of the included studies, fMRI results primarily linked walking-related sensorimotor stimulation to activation of S1/M1, SMA and Cb. These observations are not entirely aligned with those of previous neuroimaging studies using mental imagery of gait [23, 52,53,54]. Uniquely, Labriffe et al., directly compared the activation of supraspinal regions in response to mental imagery of gait and gait-related foot stimulation in a group of healthy younger adults. It was observed that mechanical stimulation, as compared to imagery, induced significantly greater activation in multiple regions (e.g., bilateral S1/M1 (especially areas related to lower limb), insula and Cb). These observations thus highlight the value of implementing MRI-compatible devices to apply walking-related stimuli to the lower extremities that mimics real walking, which can provide novel insights into the supraspinal sensorimotor control of gait that cannot be otherwise fully captured. Moreover, the stimulation and/or movements created by the different types of devices included in this review (e.g., foot sole somatosensory stimuli, stepping, pedaling, etc.), often induced different patterns in terms of negative and positive task associations, as well as recruitment of different brain regions. Therefore, in future studies, the design and characteristics of stimuli type appears to be critical and should be carefully taken into consideration.

The level of neuroimaging evidence of the activation of the brain in response to the stimulation was relatively low due to small sample sizes of participants and high heterogeneity of participant characteristics within and across studies. For example, several pilot studies focused primarily upon healthy younger adults or those with poorly-defined or under-reported inclusion and exclusion criteria. On the other hand, available evidence suggests that it is in fact feasible to simulate different sensory and/or motor aspects of walking within the MRI scanner without interfering substantially with imaging quality. Additionally, some important details were missing in several of the included publications, such as the utilization of data processing techniques, which is critical to the fMRI results. For example, eight of the included studies did not report if they completed the removal of head motion artifact [28, 37, 38, 41,42,43,44,45].

Conclusion

In conclusion, this review comprehensively summarized the current research and development achievements of the MRI-compatible devices that simulate and provide the walking-related sensorimotor stimulation when people lie motionless during the MRI scan, and thus enable the characterization of the supraspinal sensorimotor control of walking (e.g., brain’s responsiveness) via fMRI. The evidence provided by these publications suggest that it is promising to implement these devices in human study to help characterize the supraspinal control of walking. However, the small sample size of participants and varied study design limited the power of evidence. Therefore, well-powered studies with more rigorous study protocols are warranted to confirm the feasibility of implementing these devices in different populations and those preliminary observations reported here, facilitate comparison between studies, and ultimately, elucidate the supraspinal networks involved in control pertaining to the regulation of gait in relatively-healthy cohorts and in those suffering from age- or disease-related gait disorders [55, 56].

Data availability

The dataset supporting the conclusions of this article is included within the article.

Abbreviations

MRI:

Magnetic Resonance Imaging

fMRI:

functional Magnetic Resonance Imaging

Bra.Di.P.O.:

Brain Discovery Pneumatic Orthosis

pseudogait-MRCD:

Pseudogait-Magnetic Resonance Compatible Device

MARCOS:

Magnetic Resonance Compatible Stepper

LOMS:

Lower-extremity Motion Simulator

PVC:

Polyvinyl Chloride

SNR:

Signal-to-Noise Ratio

SFNR:

Signal-to-Fluctuation-Noise Ratio

S1/M1:

Primary Sensorimotor Cortex

S2:

Secondary Sensory Cortex

SMA:

Supplementary Motor Area

PMC:

Pre-motor Cortex

PFC:

Prefrontal Cortex

Cb:

Cerebellum

References

  1. Mirelman A, Bonato P, Camicioli R, Ellis TD, Giladi N, Hamilton JL, et al. Gait impairments in Parkinson’s disease. Lancet Neurol. 2019;18:697–708.

    Article  PubMed  Google Scholar 

  2. Finger CE, Moreno-Gonzalez I, Gutierrez A, Moruno-Manchon JF, McCullough LD. Age-related immune alterations and cerebrovascular inflammation. Mol Psychiatry. 2022;27(2):803–18.

    Article  CAS  PubMed  Google Scholar 

  3. Ogawa EF, Shi L, Bean JF, Hausdorff JM, Dong Z, Manor B, et al. Chronic Pain characteristics and gait in older adults: the MOBILIZE Boston Study II. Arch Phys Med Rehabil. 2020;101:418–25.

    Article  PubMed  Google Scholar 

  4. Wilson J, Alcock L, Yarnall AJ, Lord S, Lawson RA, Morris R, et al. Gait Progression over 6 years in Parkinson’s Disease: effects of Age, Medication, and Pathology. Front Aging Neurosci. 2020;12:577435.

    Article  PubMed Central  PubMed  Google Scholar 

  5. Jayakody O, Breslin M, Ayers E, Verghese J, Barzilai N, Milman S, et al. Relative trajectories of Gait and Cognitive decline in aging. J Gerontol Biol Sci Med Sci. 2021;77:1230–8.

    Article  Google Scholar 

  6. Newman AB, Simonsick EM, Naydeck BL, Boudreau RM, Kritchevsky SB, Nevitt MC, et al. Association of long-distance corridor walk performance with mortality, cardiovascular disease, mobility limitation, and disability. JAMA. 2006;295:2018–26.

    Article  CAS  PubMed  Google Scholar 

  7. Verghese J, Holtzer R, Lipton RB, Wang C. Quantitative gait markers and incident fall risk in older adults. J Gerontol Biol Sci Med Sci. 2009;64:896–901.

    Article  Google Scholar 

  8. Grillner S. Biological pattern generation: the cellular and computational logic of networks in motion. Neuron. 2006;52:751–66.

    Article  CAS  PubMed  Google Scholar 

  9. Le Ray D, Juvin L, Ryczko D, Dubuc R. Chapter 4–supraspinal control of locomotion: the mesencephalic locomotor region. Prog Brain Res. 2011;188:51–70.

    Article  PubMed  Google Scholar 

  10. Rossignol S, Dubuc R, Gossard J-P. Dynamic sensorimotor interactions in locomotion. Physiol Rev. 2006;86:89–154.

    Article  PubMed  Google Scholar 

  11. Mele G, Cavaliere C, Alfano V, Orsini M, Salvatore M, Aiello M. Simultaneous EEG-fMRI for Functional Neurological Assessment. Front Neurol. 2019;10:848.

    Article  PubMed Central  PubMed  Google Scholar 

  12. Jeannerod M, Decety J. Mental motor imagery: a window into the representational stages of action. Curr Opin Neurobiol. 1995;5:727–32.

    Article  CAS  PubMed  Google Scholar 

  13. Jeannerod M. Neural simulation of action: a unifying mechanism for motor cognition. NeuroImage. 2001;14:103–9.

    Article  Google Scholar 

  14. de Lange FP, Roelofs K, Toni I. Motor imagery: a window into the mechanisms and alterations of the motor system. Cortex. 2008;44:494–506.

    Article  PubMed  Google Scholar 

  15. Marchal V, Sellers J, Pélégrini-Issac M, Galléa C, Bertasi E, Valabrègue R, et al. Deep brain activation patterns involved in virtual gait without and with a doorway: an fMRI study. PLoS ONE. 2019;14:e0223494.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  16. Park JL, Dudchenko PA, Donaldson DI. Navigation in Real-World environments: New opportunities afforded by advances in Mobile Brain Imaging. Front Hum Neurosci. 2018;12:361.

    Article  CAS  PubMed Central  PubMed  Google Scholar 

  17. Taube JS, Valerio S, Yoder RM. Is navigation in virtual reality with FMRI really navigation? J Cogn Neurosci. 2013;25:1008–19.

    Article  PubMed Central  PubMed  Google Scholar 

  18. Diersch N, Wolbers T. The potential of virtual reality for spatial navigation research across the adult lifespan. J Exp Biol. 2019;222:jeb187252.

    Article  PubMed  Google Scholar 

  19. Jeannerod M. The representing brain: neural correlates of motor intention and imagery. Behav Brain Sci. 1994;17:187–245.

    Article  Google Scholar 

  20. Hao Y, Manor B, Liu J, Zhang K, Chai Y, Lipsitz L, et al. Novel MRI-compatible tactile stimulator for cortical mapping of foot sole pressure stimuli with fMRI. Magn Reson Med. 2013;69:1194–9.

    Article  PubMed  Google Scholar 

  21. Zhang T, Zhang K, Zhou J, Chai Y, Long Y, Wang X, et al. An MRI-Compatible Foot-Sole Stimulation System enabling characterization of the Brain response to walking-related tactile stimuli. Front Neurosci. 2019;13:1075.

    Article  PubMed Central  PubMed  Google Scholar 

  22. Kremneva EI, Chernikova LA, Konovalov RN, Krotenkova MV, Saenko IV, Kozlovskaya IB. Activation of the Sensorimotor Cortex using a device for mechanical stimulation of the Plantar support zones. 2012;38:7.

  23. Labriffe M, Annweiler C, Amirova LE, Gauquelin-Koch G, Ter Minassian A, Leiber L-M, et al. Brain activity during Mental Imagery of Gait Versus Gait-Like Plantar Stimulation: a novel combined functional MRI paradigm to Better Understand Cerebral Gait Control. Front Hum Neurosci. 2017;11:106.

    Article  PubMed Central  PubMed  Google Scholar 

  24. Jeanvoine H, Labriffe M, Tannou T, Navasiolava N, Ter Minassian A, Girot J-B, et al. Specific age-correlated activation of top hierarchical motor control areas during gait-like plantar stimulation: an fMRI study. Hum Brain Mapp. 2022;43:833–43.

    Article  PubMed  Google Scholar 

  25. Trinastic JP, Kautz SA, McGregor K, Gregory C, Bowden M, Benjamin MB, et al. An fMRI study of the differences in brain activity during active ankle Dorsiflexion and Plantarflexion. Brain Imaging Behav. 2010;4:121–31.

    Article  PubMed  Google Scholar 

  26. Noble JW, Eng JJ, Boyd LA. Bilateral motor tasks involve more brain regions and higher neural activation than unilateral tasks: an fMRI study. Exp Brain Res. 2014;232:2785–95.

    Article  PubMed Central  PubMed  Google Scholar 

  27. Belforte G, Eula G, Quaglia G, Appendino S, Cauda F, Sacco K. MR compatible device for active and passive foot movements. Int J Mech Control. 2010;11:29–38.

    Google Scholar 

  28. Belforte G, Eula G. Design of an active-passive device for human ankle movement during functional magnetic resonance imaging analysis. Proc Institution Mech Eng Part H J Eng Med. 2012;226:21–32.

    Article  Google Scholar 

  29. Doolittle JD, Downey RJ, Imperatore JP, Dowdle LT, Lench DH, McLeod J, et al. Evaluating a novel MR-compatible foot pedal device for unipedal and bipedal motion: test-retest reliability of evoked brain activity. Hum Brain Mapp. 2021;42:128–38.

    Article  PubMed  Google Scholar 

  30. Newton JM, Dong Y, Hidler J, Plummer-D’Amato P, Marehbian J, Albistegui-Dubois RM, et al. Reliable assessment of lower limb motor representations with fMRI: use of a novel MR compatible device for real-time monitoring of ankle, knee and hip torques. NeuroImage. 2008;43:136–46.

    Article  PubMed  Google Scholar 

  31. Mehta JP, Verber MD, Wieser JA, Schmit BD, Schindler-Ivens SM. A novel technique for examining human brain activity associated with pedaling using fMRI. J Neurosci Methods. 2009;179:230–9.

    Article  PubMed  Google Scholar 

  32. Mehta JP, Verber MD, Wieser JA, Schmit BD, Schindler-Ivens SM. The Effect of Movement Rate and Complexity on Functional Magnetic Resonance Signal Change during pedaling. Motor Control. 2012;16:158–75.

    Article  PubMed  Google Scholar 

  33. Promjunyakul N, Schmit BD, Schindler-Ivens SM. A novel fMRI paradigm suggests that pedaling-related brain activation is altered after stroke. Front Hum Neurosci. 2015;9:324.

    Article  PubMed Central  PubMed  Google Scholar 

  34. Martinez M, Villagra F, Loayza F, Vidorreta M, Arrondo G, Luis E, et al. MRI-compatible device for examining brain activation related to stepping. IEEE Trans Med Imaging. 2014;33:1044–53.

    Article  PubMed  Google Scholar 

  35. Martínez M, Valencia M, Vidorreta M, Luis EO, Castellanos G, Villagra F, et al. Trade-off between frequency and precision during stepping movements: Kinematic and BOLD brain activation patterns: stepping movements frequency and Precision. Hum Brain Mapp. 2016;37:1722–37.

    Article  PubMed Central  PubMed  Google Scholar 

  36. Toyomura A, Yokosawa K, Shimojo A, Fujii T, Kuriki S. Turning a cylindrical treadmill with feet: an MR-compatible device for assessment of the neural correlates of lower-limb movement. J Neurosci Methods. 2018;307:14–22.

    Article  Google Scholar 

  37. Hollnagel C, Brügger M, Vallery H, Wolf P, Dietz V, Kollias S, et al. Brain activity during stepping: a novel MRI-compatible device. J Neurosci Methods. 2011;201:124–30.

    Article  PubMed  Google Scholar 

  38. Hollnagel C, Vallery H, Schädler R, López IG-L, Jaeger L, Wolf P, et al. Non-linear adaptive controllers for an over-actuated pneumatic MR-compatible stepper. Med Biol Eng Comput. 2013;51:799–809.

    Article  PubMed  Google Scholar 

  39. Jaeger L, Marchal-Crespo L, Wolf P, Riener R, Michels L, Kollias S. Brain activation associated with active and passive lower limb stepping. Front Hum Neurosci. 2014;8:828.

    Article  PubMed Central  PubMed  Google Scholar 

  40. Jaeger L, Marchal-Crespo L, Wolf P, Riener R, Kollias S, Michels L. Test-retest reliability of fMRI experiments during robot-assisted active and passive stepping. J Neuroeng Rehabil. 2015;12:102.

    Article  PubMed Central  PubMed  Google Scholar 

  41. Jaeger L, Marchal-Crespo L, Wolf P, Luft AR, Riener R, Michels L, et al. On the modulation of brain activation during simulated weight bearing in Supine Gait-Like Stepping. Brain Topogr. 2016;29:193–205.

    Article  PubMed  Google Scholar 

  42. Takahiro I, Akira M, Kosaku S, Yasuhisa H, Yoshiyuki S. Compatibility test on lower-extremity motion simulator to fMRI. 2011 IEEE International Conference on Robotics and Biomimetics [Internet]. Karon Beach, Thailand: IEEE; 2011 [cited 2022 Aug 28]. p. 56–61. Available from: http://ieeexplore.ieee.org/document/6181262/.

  43. Ikeda T, Matsushita A, Saotome K, Hasegawa Y, Sankai Y. Pilot study of floor-reactive-force generator mounted on MRI compatible lower-extremity motion simulator. 2012. p. 311–6. Available from.

  44. Takahiro I, Akira M, Kousaku S, Yasuhisa H, Yoshiyuki S. Preliminary report of brain activities during active and passive gait-like motion. 2013 6th International IEEE/EMBS Conference on Neural Engineering (NER) [Internet]. San Diego, CA, USA: IEEE; 2013 [cited 2022 Sep 14]. p. 1566–9. Available from: http://ieeexplore.ieee.org/document/6696246/.

  45. Ikeda T, Matsushita A, Saotome K, Hasegawa Y, Matsumura A, Sankai Y et al. MRI compatibility of lower-extremity motion simulator: LoMS. 2015 IEEE International Conference on Robotics and Automation (ICRA) [Internet]. Seattle, WA, USA: IEEE; 2015 [cited 2022 Aug 29]. p. 586–91. Available from: http://ieeexplore.ieee.org/document/7139238/.

  46. Zhang Y, Meng T, Hou Y, Pan Y, Hu Y. Interpersonal brain synchronization associated with working alliance during psychological counseling. Psychiatry Res Neuroimaging. 2018;282:103–9.

    Article  PubMed  Google Scholar 

  47. Travassos C, Sayal A, Direito B, Pereira J, Sousa T, Castelo-Branco M. Assessing MR-compatibility of somatosensory stimulation devices: A systematic review on testing methodologies. Frontiers in Neuroscience [Internet]. 2023 [cited 2023 Sep 22];17. Available from: https://www.frontiersin.org/articles/https://doi.org/10.3389/fnins.2023.1071749.

  48. Yang B, Tan U-X, McMillan A, Gullapalli R, Desai JP. Design and control of a 1-DOF MRI compatible pneumatically actuated Robot with Long Transmission lines. IEEE ASME Trans Mechatron. 2011;16:1040–8.

    Article  PubMed Central  PubMed  Google Scholar 

  49. Boland BL, Xu S, Wood B, Tse ZTH. High speed pneumatic Stepper Motor for MRI applications. Ann Biomed Eng. 2019;47:826–35.

    Article  PubMed  Google Scholar 

  50. Perrachione T, Ghosh S. Optimized design and analysis of sparse-sampling fMRI experiments. Front NeuroSci. 2013;7:55.

    Article  PubMed Central  PubMed  Google Scholar 

  51. Field AS, Yen YF, Burdette JH, Elster AD. False cerebral activation on BOLD functional MR images: study of low-amplitude motion weakly correlated to stimulus. AJNR Am J Neuroradiol. 2000;21:1388–96.

    CAS  PubMed Central  PubMed  Google Scholar 

  52. Hamacher D, Herold F, Wiegel P, Hamacher D, Schega L. Brain activity during walking: a systematic review. Neurosci Biobehavioral Reviews. 2015;57:310–27.

    Article  Google Scholar 

  53. Jahn K, Deutschländer A, Stephan T, Strupp M, Wiesmann M, Brandt T. Brain activation patterns during imagined stance and locomotion in functional magnetic resonance imaging. NeuroImage. 2004;22:1722–31.

    Article  PubMed  Google Scholar 

  54. la Fougère C, Zwergal A, Rominger A, Förster S, Fesl G, Dieterich M, et al. Real versus imagined locomotion: a [18F]-FDG PET-fMRI comparison. NeuroImage. 2010;50:1589–98.

    Article  PubMed  Google Scholar 

  55. Haller S, Bartsch AJ. Pitfalls in FMRI. Eur Radiol. 2009;19:2689–706.

    Article  PubMed  Google Scholar 

  56. Oakes TR, Johnstone T, Ores Walsh KS, Greischar LL, Alexander AL, Fox AS, et al. Comparison of fMRI motion correction software tools. NeuroImage. 2005;28:529–43.

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

Dr. Junhong Zhou is supported by a grant from the National Institutes on Aging (1K01AG075180-01).

Author information

Authors and Affiliations

Authors

Contributions

YH contributed to the literature search, data analysis, data interpretation, and preparation of this manuscript. YH and JZ contributed to the study design, data analysis, data interpretation, and manuscript preparation. DB, BM and JZ contributed to data interpretation and revising of the manuscript. All authors approved the submission of the manuscript.

Corresponding author

Correspondence to Dapeng Bao.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hong, Y., Bao, D., Manor, B. et al. Characterizing the supraspinal sensorimotor control of walking using MRI-compatible system: a systematic review. J NeuroEngineering Rehabil 21, 34 (2024). https://doi.org/10.1186/s12984-024-01323-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12984-024-01323-y

Keywords