EMG-driven control in lower limb prostheses: a topic-based systematic review

Abstract

Background

The inability of users to directly and intuitively control their state-of-the-art commercial prosthesis contributes to a low device acceptance rate. Since Electromyography (EMG)-based control has the potential to address those inabilities, research has flourished on investigating its incorporation in microprocessor-controlled lower limb prostheses (MLLPs). However, despite the proposed benefits of doing so, there is no clear explanation regarding the absence of a commercial product, in contrast to their upper limb counterparts.

Objective and methodologies

This manuscript aims to provide a comparative overview of EMG-driven control methods for MLLPs, to identify their prospects and limitations, and to formulate suggestions on future research and development. This is done by systematically reviewing academical studies on EMG MLLPs. In particular, this review is structured by considering four major topics: (1) type of neuro-control, which discusses methods that allow the nervous system to control prosthetic devices through the muscles; (2) type of EMG-driven controllers, which defines the different classes of EMG controllers proposed in the literature; (3) type of neural input and processing, which describes how EMG-driven controllers are implemented; (4) type of performance assessment, which reports the performance of the current state of the art controllers.

Results and conclusions

The obtained results show that the lack of quantitative and standardized measures hinders the possibility to analytically compare the performances of different EMG-driven controllers. In relation to this issue, the real efficacy of EMG-driven controllers for MLLPs have yet to be validated. Nevertheless, in anticipation of the development of a standardized approach for validating EMG MLLPs, the literature suggests that combining multiple neuro-controller types has the potential to develop a more seamless and reliable EMG-driven control. This solution has the promise to retain the high performance of the currently employed non-EMG-driven controllers for rhythmic activities such as walking, whilst improving the performance of volitional activities such as task switching or non-repetitive movements. Although EMG-driven controllers suffer from many drawbacks, such as high sensitivity to noise, recent progress in invasive neural interfaces for prosthetic control (bionics) will allow to build a more reliable connection between the user and the MLLPs. Therefore, advancements in powered MLLPs with integrated EMG-driven control have the potential to strongly reduce the effects of psychosomatic conditions and musculoskeletal degenerative pathologies that are currently affecting lower limb amputees.

Introduction

Microprocessor-controlled lower limb prostheses

Modern Microprocessor-controlled lower limb prostheses (MLLPs) represent a class of prosthetic devices that can simulate the joint’s biological behavior through real-time adaptive control driven by the sensory information acquired from embedded sensors (e.g. encoders, load and force cells) [42, 43]. Modern advancements in actuation and electronics have primarily led to the development of variable damping (passive) MLLPs and then to powered (active) MLLPs [131]. The main difference is that the former class of devices can only change the joint impedance, but unlike the latter, cannot actively generate net positive power. Compared to their fully passive (non-microprocessor controlled) predecessors, both types of devices can reproduce a broader repertoire of the human behaviour by restoring more dynamic functionalities.

In particular, variable damping lower limb prostheses guarantee the restoration of almost healthy-like locomotion of energy neutral or dissipating actions, such as walking, stair descent and sitting down. However, users have no direct control over the device [42]: it is the device itself that decides how to behave based on the sensors recorded information. This condition tends to lead to high cognitive fatigue and excessive energy consumption, especially during more complex activities [84]. User perception of inadequate controllability of the device, specifically a lack of intuitive control, reduces the acceptance rate of the lower limb prosthesis, which consequently leads to its abandonment [21, 108]. While comfort remains one of the key factors in prosthesis rejection [9, 99, 112], poor mobility is one of the leading causes of eventual device abandonment [44, 46, 104]. Additionally, users that do not abandon their prostheses incur high risks of developing a series of neuromusculoskeletal disorders and cardiovascular diseases [12, 45, 75, 92]. This is related to the inability of variable damping MLLPs to provide positive power: most types of basic healthy-like locomotion (e.g. walking) are governed by phases of positive power output [60, 88, 89, 95]. As a consequence, compensatory movements and gait asymmetry increase biomechanical stresses, particularly on the healthy biological joints of the amputee, causing articular pain to the knee, hip and back [8].

Active powered prostheses have been suggested to address the desire of being able to exert positive power during locomotion. In theory, these devices should allow for more natural gaits, and enable a wider range of possible movements and energy-generative actions, such as sloped gait, sit-to-stand, stair climbing and running [124]. However, their benefits have not yet been validated through biomechanical, performance-based and patient-reported metrics [42, 126]. Most likely, this validation stage has not yet been reached because of the poor user acceptance of these devices. Their poor acceptance can partly be explained by mechanical challenges that remain difficult to overcome. For example, the powered prostheses on the market are noisier, heavier, less smooth, and have shorter battery life than variable damping MLLPs. However, in addition to mechanical challenges, similarly to the variable damping prostheses, most powered prostheses do not provide direct control to the end-user: their control relies solely on interpreting embedded sensor data. Possibly, lack thereof is even more critical if the device allows for a wider range of activities.

Neuro-control architecture

Concurrently, trends in prosthetics, orthotics and Human–Robot Interaction (HRI) are pushing not only for the restoration of essential human locomotion, but also for the user to have direct control to the device through the neural pathways [52]. Neuro-controllers in fact have the capabilities to decode the neural activity either from the central or the peripheral nervous system in order to control external devices. In particular, it is possible to use EMG sensors to measure and decode users’ motion intention directly from neural activity using muscles as terminal amplifiers of motor afferent commands [94].

Human locomotion neuro-control comprises two particular sets of commands: volitional and rhythmical. Rhythmic locomotion occurs when humans use repetitive limb movements to translate in space, such as walking and running. Usually, this class of motor patterns does not involve conscious intervention on behalf of the subject, but they result from sensory-motor reflexes activated from specific neural networks [81]. Scientific evidence locates these so-called Central-Pattern Generators (CPGs) in the spinal cord or in the brain stem and they trigger organized muscle contraction in a cyclic manner [2, 30]. Their activation and regulation can both derive from Central Nervous System (CNS) inputs and sensory-motor feedback [31, 82]. Volitional movements instead are a broad class of movements that involve motion planning and motor control, like non weight-bearing knee flexing. Conscious movements require constant and vigilant attention from the CNS involving different cortex areas and intensive collaboration across the diencephalon, brainstem, and cerebellum [73].

To achieve the same locomotion control, MLLPs have different levels in their control architecture that is structured similarly to the human neuro-control system. This particular multi-layer control framework is referred to as a hierarchical controller, and shown in Fig. 1 [124]. In particular, the high-level control is tasked with identifying and quantifying the user’s intention. This primarily regards selection of user activity mode (e.g. walking, slope ascending or sitting [111]) and, secondarily, context recognition (e.g. ambulation speed and swing or stance phase [119]). The more these selections stem from user awareness and volition, the more difficult it is to identify user volition based only on embedded sensors, such as joint encoders, load cells and an Inertial Measurement Units (IMU). However, the use of EMG sensors is a valid choice for the reliable identification of volitional intent, which paves the way for direct neuro-control of the device [56].

Fig. 1
figure 1

(adapted from [124])

Hierarchical controller schematic representation. Comparison between a generalized control framework for microprocessor-controlled lower limb prostheses (right) and the human motor control (left). The figure displays the classic interactions between the Hierarchical Controller layers and user-device-environment. Controller level-specific tasks are listed in the figure

The identified intent from the high-level controller is fed to the mid-level controller, which translates it to device state control, such as a target joint impedance, stiffness or angular trajectories. Finally, the device-specific controller in the low level interprets the output of the mid-level controller, and closes the loop with the device-specific physical hardware, e.g. by controlling motor currents. The mid- and low-level controls are well comparable to the human’s brainstem and spinal cord, which—in contrast to the high-level control—can function well by relying solely on the aforementioned system dynamics without requiring much input or intervention of the human’s awareness and volition.

Scope of the review

Among different neuro-controllers, roughly 80% of the papers collected on lower limb prosthesis control published after 2010 are related to EMG-driven control [131], which indicates the recent popularity of this subject. This particular research interest in myoelectric control—also known as EMG-driven control—is related to the necessity of novel technological solutions to support the development of new powered lower limb prostheses [50, 52, 131]. Myoelectric control represents a possible candidate for introducing a user’s direct control for powered MLLPs. Although the literature promotes this technology and its positive benefits for the end user, such as reduction of phantom limb pain [80, 96], it is not mature enough for powered MLLPs. In fact, no commercial device is available with such technology, unlike their upper limb counterparts.

Only one comprehensive analysis has been found that focuses on EMG-driven control in lower limb prostheses [41]. Whereas various reviews have been published on MLLPs, most of these focus on mechanics and control, and either do not discuss EMG-driven control [50, 52, 69], or dedicate a short section to it [42, 84, 124, 131]. Instead, whereas numerous other reviews focus specifically on myoelectric control, they either do so for generic HRI [1, 3, 93, 94, 97, 103, 109], or for upper limb prosthetics specifically [47, 102, 107].

In the attempt to differentiate from previous reviews, we are focusing our survey on the novel and more challenging implementations of EMG-driven control in high-level rather than mid- or low-level control. Information and discussions about mid- and low-level control can be found in [52, 69, 124, 131]. Therefore, this work aims at providing a detailed and organized systematic overview of myoelectric control for powered MLLPs prostheses. The main objective of this review is to analyze the merits and drawbacks of the various implemented EMG-driven controllers. To do so, we defined four major topics of investigation categorizing the most important and common characteristics in the available literature. This analysis was conducted to provide an unbiased and systematic evaluation of the literature on EMG-MLLPs. In addition, the goal is to discriminate the potentialities and limitations of each technological embodiment to define a clear path for future research investment.

Methods

Eligibility criteria

The first eligibility criteria for this review was the actual dissertation of a EMG-driven control for MLLPs. Papers were considered only if presenting: (1) an implementation of the myoelectric controller, (2) a complete description of its architecture, and (3) results on a physical or simulated device.

Since EMG-driven control for MLLPs represents a niche research sector in both HRI and prosthetics, authors decided to not include any limitation in the year of publication, type of publication (e.g., journal or conference papers) and language.

Information sources, search strategy and selection process

Searches have been conducted in several major scientific research databases. The following databases were chosen based on the number of results obtained with the search phrase “myoelectric control lower limb” in Google Scholar: PubMed, ScienceDirect, IEEE Xplore, Springer Link, ASME, NCBI and ResearchGate. Successively, combinations of the words “myoelectric”, “EMG”, “control”, “lower limb”, “knee”, “ankle” and “prosthesis” were used for the actual search in the cited databases. Results were not screened based on the date of publication and type of publication. Databases were last consulted for this review in May 2021.

Due to high prevalence in the results of papers related to upper limb prostheses, orthoses and other robotic devices, and other papers not related to EMG-driven lower limb prosthesis controllers, a manual selection was made. Only papers regarding knee and ankle prosthesis controllers driven from EMG acquisitions were included based on title and abstract. Content of the remaining articles was reviewed. Papers were included in the review process if they presented an actual implementation of the myoelectric controller, a complete description of its architecture, and results on a physical or simulated device. Finally, all cited papers from the manuscripts obtained from the screening strategy were collected and went through the same search and selection process.

The diagram in Fig. 2 illustrates the number of manuscripts that were identified, screened and finally included in this study.

Fig. 2
figure 2

Systematic review flow diagram. The PRISMA flow diagram for the systematic review, detailing the database searches, the number of abstracts screened and the full texts retrieved

Data items: topics of investigation

The data collected from this review is divided according to four major topics used to structure the literature investigation:

  • Neuro-control: This aspect discusses how the nervous system, through the muscles, can control the lower limb prosthesis. It concerns the use of input neural signals (i.e. EMG) for the generated output movement, rather than the specific implementation of controllers. In particular, it explores the types of movement that can be restored, namely either rhythmic locomotion or volitional movement. This is consequently reflected on the implementation of different EMG-driven control strategies: Computational Intrinsic Control (CIC) or Interactive Extrinsic Control (IEC) [84].

  • EMG-driven working principles: The methodologies employed in EMG-driven controllers to translate user intention and volition to high-level control parameters have been group in three major categories: direct control, pattern recognition, and model-based. Each of this class adopt a different principle to interpret user volition and intention from EMG signals and translate it in to a high-level control.

  • Neural input and processing: Depending on the type of MLLP, the level of amputation and control class, and the choice of muscles, different sensors and processing methods can be applied. Identifying these differences or similarities is essential to confront methodologies in relation with the final results.

  • Performance assessment: This topic discusses the collection of meaningful reported results in the reviewed literature to provide the reader an overview of the achieved performance of the controllers.

Data collection process, study of risk bias assessment, effect measures

The first author defined the main source databases and conducted the search and selection process. No risk bias has been identified for the data selection process. However, to avoid possible apophenia and contextual biases in the collection process, the second author revised the proposed data items and defined detailed subcategories for each item (topic of investigation). For each subcategory, the first author chose effect measures that were recurrent among all reviewed manuscripts. Effect measures are collected in different form depending from the topic of investigation. The rest of the authors inspected the synthesis results to ensure no incongruity among the collected measures and the data clustering.

Synthesis methods

The authors chose to synthesize the collected data in a tabular form: one table for each investigated topic. Data was catalogued in sub-fields representing relevant information for the implementation and evaluation of the EMG-driven controller in MLLPs. In each table, all the collected papers are cited and grouped by the type of working principle. Paper citations for each working principle are ordered by date of publication. If manuscripts employ identical concepts, they are grouped in a single table entry and cited collectively.

In addition, a graphical diagram was designed to navigate the collected data in the tables and to show the possible interconnections between the four topics of investigation (Fig. 3). The diagram organizes and synthesizes the collected literature according to the obtained data items. Some of the most representative papers were moreover used as examples for better displaying the obtained results.

Fig. 3
figure 3

Graphical overview of the reviewed literature, grouped by the same four topics used in the manuscript. The references are grouped by application (knee, ankle or both), the EMG-driven control working principle and type of neuro-control (type of movement restored). Some of the most relevant works are detailed additionally by listing the employed additional inputs, number of EMG channels and controller delays, if known. Five classes of additional data inputs were recognized (apart from EMG signals). Motion Capture (MoCap) and sensorized treadmills with force platforms are used to acquire body kinematic and dynamic data; IMU sensors are usually integrated to obtain orientation of the lower limb segments; footswitch and loadcells are installed to acquire the force exchanged with the environment; encoders and goniometers are used to measure joint angles; finally, visual or haptic feedback is sometimes provided to the user to encourage a correct employment of the device. On the lower part of the graph, the most used measurements for each control class validation is displayed. NS not stated; referenced work belong to a hybrid model-based and pattern recognition control class, see “EMG-driven working principles” section

Results

This section presents the complete listing of the review findings, divided in the four topics of investigation as described in “Data items: topics of investigation” section.

In order to aid the reader in navigating the results and to give a better understanding of the interconnection of these four topics, a graphical overview of the reviewed literature is presented in Fig. 3. From this graph it is possible to advance some preliminary results. For example, looking at neuro-control, the number of works that investigate the control of prosthetics in volitional movement (IEC) and rhythmic locomotion (CIC) is balanced (29 papers each), but pattern recognition controllers are more numerous in literature. It can be additionally noticed that there is no particular preference regarding the control type with respect of the type of prosthetic joint. A similar result has been found for the type of additional sensors, when inspecting the possible control signal inputs. Instead, works that employ pattern recognition principles generally use more EMG channels (\(6.6\pm 3.1\) channels on average) than those employing direct control and model-based control principles (respectively \(3.2\pm 1.9\) and \(2.5\pm 1.1\) channels on average). Volitional movement controllers use generically more often visual/haptic feedback (almost 50% of the collected papers). Looking at the performance assessment, no significant pattern can be determined on the control delay, which refers to the time between muscle excitement and control execution. It is anyway interesting to notice that two out of ten of the selected papers did not report any information about this parameter, while being an important metric to evaluate the HRI. Finally, the diagram shows a strong preferences for performing biomechanical evaluations (joint angle and torque trajectories) on IEC controllers, while perform-based measures (locomotion classification accuracy) are preferred instead on CIC.

Specific results of each of the four categories and their interconnections are listed separately in the following sections.

Neuro-control

Results on the neuro-control abilities are collected in Table 1.

Table 1 Overview of the neuro-control capabilities of the device

Martin et al. categorized rhythmical and volitional type of movements in two general behaviours that a high-level controller can implement: CIC and IEC [84]. These essentially distinguish between controllers that are based on the conscious involvement of the user (volitional or IEC) and those that are not (rythmical or CIC) (Fig. 4). Correspondingly, the adoption of this particular subdivision allows us to classify the EMG-driven controllers based on how the neural information is employed (Fig. 3).

Fig. 4
figure 4

Computational Intrinsic Control (CIC) and Interactive extrinsic Control (IEC) approaches for lower limb prosthesis neuro-control. Comparison between CIC (on the left) and IEC (on the right) from motor task generation to EMG-driven control. A CICs chose the correct control among the states implemented in the control board based on the EMG signals generated during a rhythmic locomotion; B IECs instead transform EMG recorded patterns to a specific continue modulation of the prosthetic joint

CICs typically decode the user’s motor intention and prosthetic device state through data from embedded neural sensors This information is then employed to establish an appropriate control in order to accommodate for the changes in the locomotion [121]. This is observable in Table 1: the majority of CICs are characterized by a neural-control based on adaptive controls depending on locomotion recognition. CICs represent the state-of-the-art in the control of non-EMG MLLPs and they have been explored with different approaches, both in the literature and in commercially available devices [42].

Instead, IECs are designed to guarantee continuous communication between the user and the device so as to directly modulate the prosthesis’ state (Table 1). In fact, this type of control, differently from CIC, cannot be implemented in non-EMG MLLPs since no neural input is provided to the prosthetic device. As such, IECs have enabled the control of lower limb movements that were not possible with CIC.

From the obtained results, no particular correlation among the remaining fields was found. This suggests that there is no particular influence of the choice of neuro-control on the specific prosthetic device, but only on the prioritized type of movement to be restored.

EMG-driven working principles

Table 2 provides an overview of the control strategies and working principles of the reviewed EMG-driven approaches for MLLPs. The table also shows that the reviewed literature has been classified according to the following three types of control: Direct Control, Pattern Recognition Control and Model-Based Control.

Table 2 Overview of lower limb EMG-driven controllers working principles

Independently from the type of controller, an essential feature is the walking control. Walking represents the most important movement in bipedal locomotion and is therefore essential to be restored in lower limb amputees. Consequently, the primary use of EMG signals is to accomplish healthy-like ambulation. The use of EMG signals and the design of the controller depends on the type of EMG-driven controller.

Direct EMG-driven controllers refer to all those controllers that employ EMG as input of a specific function \(y = f(x_{\text {emg}})\), where y represents the variables to be controlled during the walking cycle for the generation of the signal reference, such as joint angle or torque (Fig. 5a). Examples of these are proportional controllers [29, 64], regression functions [24, 37] or mapping-transformation functions between processed multi-channel EMG acquisitions and control parameters [56, 57, 62].

Fig. 5
figure 5

Graphical representation of the three possible solutions for EMG-driven control. EMG signals are acquired from the stump of the amputee and used as input for the high-level controller, depending on how the neural signals are used three type of EMG-driven controllers can be implemented: A direct control, B pattern recognition, C model-based; the control signal output is then used from the internal lower level control of the prosthetic device

Pattern recognition control represents the largest class. They are reminiscent of (FSM)-based controllers, which are currently used in commercial MLLPs (Fig. 5b). FSM-based controllers distinguish a number of control states (actions) through state classification (perception), a concept similar to that governed by pattern recognition algorithms, a branch of Machine Learning (ML) [18]. Typical pattern recognition algorithms identify particular signals’ signatures (features) that can be observed under particular conditions. These features can be used, consequently, as transition rules in the controller to distinguish the different working conditions [70, 98, 106], such as slope adaptation (SlA) and speed adaptation (SpA). For this reason, as illustrated in Table 2, pattern recognition control algorithms provide only a pre-defined number of locomotion types with respect to the other controllers, such as only for stair ascend and descend, standing and sitting. Instead, direct controllers allow theoretically all types of motion.

The last class of EMG-driven controllers employ Musculoskeletal (MS) models to reproduce human kinematics and dynamics in locomotion [7, 83, 119] (Fig. 5c). Modeling in bio-engineering research has been used to study biological systems, whose characteristics are not directly measurable. In case of lower limb prostheses, it is possible to use MS to estimate joint impedance starting from the measured external forces and joint trajectories [34]. Additionally, with the use of EMG signals and translating them to appropriate muscles activation, these models can be used to modulate voluntarily tracked joint impedance [6, 61, 130, 136].

Another important distinction regarding the EMG-driven working principles is the different partitioning between user-training and control-calibration time. Both describe the time the user needs to spend with the prosthesis to achieve an effective Human-In-The- Loop (HITL) control of the device [79]. In fact, intra- and inter-subject variability of the biological system and of the signals involved in the control loop increase uncertainty in the control performance [132]. Stochastic variability within HITL can be reduced by tuning internal control variables, but it cannot be removed completely [133]. For this reason, the user has to be trained to avoid control errors and understand how to use the device optimally. As such, while calibration time is correlated to the number of tunable variables and provides an indication of control complexity, the training time is an indicator of how seamless and intuitive the control is. From the collected data (Table 2), it can be observed that the ratio between calibration and training time changes drastically in pattern recognition control. In fact, multiple recordings are necessary for the appropriate calibration of the FSM, but after this initial effort the user needs a minimal amount of training. The contrary can be said for the direct control: short initial parameter tuning, usually based on previous analysis, followed by prolonged training sessions.

The classification of EMG-driven control type is closely related to the neuro-control strategy since the analyzed literature shows that the pattern recognition algorithms are all CIC controllers, whereas direct control and model-based approaches are IECs. This can be observed also in Fig. 3. While the neuro-control distinction allows the reader to understand which are the target motor tasks intended to be restored, the working principle describes the methodology employed for the control.

Neural input and processing

Table 3 lists the used muscles and applied signal processing techniques to provide input data to the EMG-driven controller.

Table 3 Overview of the used input signals and applied processing

Assuming that a robotic prosthesis can be fastened to the user stump and replace the missing limb, it may be possible to embed the EMG acquisition system on that device. Surface EMG activity is characterized by a high temporal resolution, which makes it practical for the control of an external device. However, signal non-stationarity, motion artifacts, electrode-skin conductivity variations, and channel cross-talk require computationally-demanding signal processing techniques [74]. In order to mitigate these problems, additional mechanical sensors, such as load cells (see also Fig. 3), can be introduced to strengthen the reliability of the control [66]. The sensors listed in Table 3 refer only to those used for the actual control loop: either as input for the device’s high-level control or for its calibration.

Even when using extra sensors, proper muscle choice and EMG processing (e.g. filtering) are essential to guarantee the robustness of the controller.

In order to decode the user’s movement intention, a layer of feature extraction and classification can be added to the control system. Useful features for EMG control are well described in literature [74]. Nonetheless, the choice of the sampling window length is often tailored to the specific processing choices. In fact, the size of the window has to be as large as possible to guarantee the stability of the extracted features, but at the same time—due to signal non-stationarity—the length is calibrated to meet the quasi-stationarity condition.

Performance assessment

Results concerning control validation and reliability are listed in Table 4, which collects the main findings presented in the selected references.

Table 4 Overview of the evaluation measurements and related results

The efficacy of an EMG-driven control framework is always a trade-off between intuitiveness, system response time and accuracy of movement selection [35]. Consequently, it is challenging to define a threshold of acceptability of a given implemented solution. Bias and variance of the myoelectric control parameters, as well as in the input signals, depend on numerous conditions and they can heavily impact performance [3]. The reviewed manuscripts showed limited amount of evidence on assessing patient safety and security, probably due to their prototypical stages of the development. However, there are studies focusing exclusively on EMG-driven control faults and safety issues, which analyze critical situations for stumbling and falls [145].

The different number of subjects and the average results presented in Table 4 show that a standardized evaluation is missing for defining the overall performances of MLLP EMG-driven controllers, even at the prototype stage.

During this review, the authors found a strong correlation between the type of evaluation and the type of implemented neuro-control. In fact, while IEC manuscripts focus their analyses on biomechanical measurements, CIC controllers focus mainly on evaluating performance-based measurements.

An analysis of the literature also shows that control delay is an important parameter to evaluate the controller performance in real scenarios [137]. In particular, to guarantee control stability while considering the HITL factor, the total maximum control delay between the signal window sampling and the actual output value has to be less than the human physiological electromechanical delay \(d_\text {EM}\). The human \(d_\text {EM}\) is caused by the time that a neural signal requires to generate the electrical depolarization of the muscle tissue, to consequently result in a mechanical force, and finally joint displacement. The delay has been estimated to be \(d_\text {EM} \approx 100\) to \({150}\;{\text {ms}}\) for the lower limb [123, 129]. If the generation of the prosthetic mechanical movement requires longer time than \(d_\text {EM}\), the prosthesis will be always delayed with respect to the volition of the user, which limits the usability of the controller and may generate risky conditions. Even though the prosthesis’ motion generation delay is a crucial metric to evaluate the EMG-driven control, close to half of the reviewed manuscripts do not provide such information. Long delays are especially problematic if they are prone to a logical deadlock, where the system comes to a halt because its subsystems (human and the prosthesis) are waiting for each other to take action. For example, if the controller requires completion of a (partial) step cycle to detect a user-intended switch from walking to stair climbing, while the user is not at comfort or able to perform the first stair climbing step with the presently selected walking controller, then the switch will never occur.

Discussions

The results presented in “Results” section can be further discussed to identify patterns, prospects and limitations of the various approaches, to understand how future research and development into EMG-driven controlled MLLPs is most promising.

The structure of the discussion follows that of the results, i.e., uses the same four topics as subsections. Furthermore, this section presents an additional discussion in “Clinical considerations” section, regarding the clinical implications of developing an effective neuro-controller, in particular an EMG-driven controller. This is important to provide a general interpretation of the results in the context of other evidence.

Neuro-control

High-level controllers for lower limb prostheses, and in general lower limb robotics, aim to restore both rhythmic locomotion (CIC) and volitional movements (IEC). Unfortunately, very little is known about how these two processes cooperate in humans, but it has been proposed that these two paths can bidirectionally interact and cooperate with each other [20]. This lack of knowledge generates a particular dichotomy of EMG-driven control in restoring either rhythmic or volitional movement. CIC and IEC aim exactly in the control of either type of movements (Fig. 4).

Similarly as in human CPGs, prosthetic CIC does not require any conscious human involvement [138]. Rhythmic motor patterns, like walking, compose almost the totality of human basic locomotion, and therefore constitute the most important type of movements to restore in amputees. Rhythmic motor patterns are easily detectable and reproducible, by employing basic concepts of data analysis on prosthesis kinetic and kinematic measurements [100, 127]. For this reason, pattern recognition control techniques are employed for implementing CICs and are the most widely used. In fact, as stated in “EMG-driven working principles” section, pattern recognition uses data from embedded sensors in the prosthetic device (e.g. load/pressure cells, foot switch, joint encoders and IMU) to identify the locomotion type and change the control law parameters accordingly. The introduction of EMGs and therefore the integration of user volition information in this control paradigm has been investigated mainly with two objectives: using the myoelectric signals as additional information to increment the number of classes of movements that can be controlled, and boosting their recognition accuracy. These controllers therefore inherited the same controller capabilities of their non-EMG MLLPs predecessors: high reliability during rhythmic locomotions in spite of their lack of intuitive control. Therefore, CICs are the most effective option when the main objective of the MLLPs is to restore walking, which is the case for a major target group of lower limb prosthesis users. In contrast to rhythmic movements, volitional motions are not characterized by phase-dependent trajectories, nor by correlated conditions between repetitive movements. Due to the redundancy in the musculoskeletal system, identical limb configurations can be generated from different muscle activations [77]. This particular property of the human musculoskeletal system makes it significantly more difficult to design a controller that is able to obtain a reliable solution through the processing of only EMG signals for the voluntary control of the artificial joint. Direct and model-based controllers in IECs tackle this issue in two different ways: the former employs signal processing to reduce these aforementioned redundancies, while the latter uses modeling in order to characterize them. Despite this technological difficulty, users can employ this type of control to voluntarily and continuously modulate joint flexion-extension with higher degree of freedom respect to the previous class of controllers.

Collected results showed that only two research groups attempted to unify CIC and IEC in a unique control strategy [72, 113]. The first research was conducted at the former Center for Bionic Medicine, Rehabilitation Institute of Chicago [58, 111]. They adopted a pattern recognition control algorithm, in combination with Targeted Muscle Reinnervation (TMR), to implement a hybrid CIC-IEC. Depending on the device state, the FSM could decide either to control rhythmical locomotion or, in case of a non-bearing motor task, give the user the freedom to control the joint voluntarily.

Another exception to the CIC/IEC dichotomy can be seen in the work conducted at the MIT Media Lab [71, 72, 130]. In their studies, the EMG signal proportionally regulates the gain of a Hill-type muscle model to generate additional plantar-flexion force. Deviation from the intrinsic controller output was enabled depending on the type of locomotion performed and the level of muscular activation: this was especially used to generate an additional push-off when required by the user. Their approach employed standard pattern recognition techniques for locomotion control, while applying direct voluntary or model-based techniques to deviate from such intrinsic control during particular locomotive states. These studies recognized the strengths of the two strategies and attempted to simultaneously preserve both high reliability and seamless control.

This particular choice on the type of neuro-control and the type of movement to be addressed has a direct consequence on the type of controller that should be implemented. Their particular working principle is discussed in the following section.

EMG-driven working principles

Walking represents the most important locomotion in humans: on average 10,000 steps per day for the younger population (\(\le 65\) years old) and 7000 for the older (\(>65\) years old) [10]. Consequently, SlA and SpA are usually the most implemented, since they are essential to accommodate different walking patterns based on the cadence and terrain conditions. Table 2 displays the additional modalities (locomotion types) that are allowed to be performed from the controllers, apart from walking. While an EMG-driven pattern recognition approach has to declare the modalities that can be controlled specifically in the framework design, direct-control and model-based approaches are able to deliver the control for any type of locomotion. However, even if they theoretically could, there is a lack of experimental validation for this aspect. In fact, some of the reviewed works did not perform any validation regarding such capabilities, as they only report results of ground-level walking [26, 37, 61].

Despite this limitation, the strength of pattern recognition algorithms, such as ML, is the ability of learning and recognizing peculiar features from a given data set. This skill becomes useful for the case of multi-dimensional and complex data streams that have to be analyzed and labeled in real time, such as EMG. Moreover, EMG signals are highly irregular and non-stationary, and amputees could develop abnormal co-contraction features in the residual limb, making techniques of direct control particularly difficult.

Another important characteristic that distinguishes pattern recognition techniques from other control techniques, is the reduced user training time, as can be seen in “EMG-driven working principles” section. Clinical evaluations of myoelectric prostheses have correlated the reduced training time to a higher device acceptance [35]. Even though there are no standards to establish an optimal training time, prolonged and repetitive sessions might result in cognitive and physical exhaustion. However, it has to be noticed that no pattern recognition study explored multiple testing sessions and therefore dealt with recalibration issues. Depending on the robustness of the developed pattern recognition EMG-driven controller, calibration time might be necessary before each session, correspondingly worsening device acceptance.

This discussion has clarified that the choice of the type of the controller affects the type of processing and the type of signals. Further discussion on the type of processing related to the type of EMG-driven working principle is discussed in the next section.

Neural input and processing

The choice of which muscle signals to use represents a fundamental topic in EMG-driven control. This choice is influenced not only by the architecture of the controller, but additionally by the artificial joint’s mechanical and electronic design, from the amputation level and finally by the surgical muscle reattachment [51]. The latter two points are subject-specific, which leads to user inter-variability regarding choices of the muscle signals and consequently the EMG-driven control technique. Additionally, the biological joints are controlled from a different number of muscles, which limits the number of signals that can be used after amputation depending on the joint. Moreover, the presence of bi-articular muscles—and therefore the kinematic and dynamic coupling between the different joints—may also affect the number of channels that can be used. Finally, MLLPs mechanics and electronics (such as prosthesis weight, volume, and battery-life) can affect the number of sensors that can be embedded.

Related to this issue, the reviewed literature tended to go in two opposite directions. Some of the studies used a small number of EMG-sensors, and placed them on big proximal muscle groups that would still be present after the amputation, as in [6, 24, 29, 56, 61, 63, 64]. Instead, other studies acquired readings from multiple muscles in order to collect redundant information, and to analyse how the controller performance changes by using a variety of signal processing techniques [66, 70, 120]. Whereas the latter requires an initial analysis, this approach can be used to better tailor the controller to user necessities, leading to an improved performance when the number of EMG sensors is higher. This particular choice is usually related to pattern recognition controllers and explains why this class of controllers has usually more EMG channels.

Additional sensors have been extensively used in EMG-driven control in order to improve their accuracy (Fig. 3). In fact, a common strategy is to add mechanical sensors (e.g. IMU and pressure cells) to the prosthetic device to have more reliable information to compensate for the high variability of EMG signals and their sensitivity to noise. For example, using a foot switch for detecting gait events (e.g. heel strike and toe off) is still the main method for the identification of stride phases. This identification can be used additionally to adapt different lower-level control strategies or for disabling/enabling the EMG-driven control with respect to standard approaches [6, 58].

The use of supplementary mechanical sensors (e.g. encoders) in addition to EMG electrodes can also be used for restoring the missing sensory feedback from the amputated limb. It was found that the most common approach was to provide the user the joint angle position during joint movements through visual representation. Indeed, feedback is particularly important in the IEC class of controllers. In this particular case, since the user is in continuous and full control of joint flexion and extension, it is beneficial to return information about the MLLPs configuration. The use of visual feedback, though, is impractical during every-day activities. A potential solution is to employ a haptic feedback, such as vibrotactile stimulus, coding for joint angle position, as suggested by [6, 62]. However, only one work was found to investigate the use of haptic feedback in conjunction with EMG-driven control [14]. Their results suggested that this was mostly beneficial when visual feedback was limited.

Processing EMG signals usually follows two possible directions, which is mainly influenced by the type of neuro-control implemented from the high-level controller. For IEC, the common approach is to extract the fully rectified envelope of the signals around the most informative bandwidth (around 10 to 500 Hz), and then to normalize the relevant signals for their maximum values. Instead, for the case of CIC, where frequency-based features are usually employed, band-pass filtering is applied in order to delete low- and high-frequency noise components.

Classification of EMG features, as shown in Table 3, is usually applied in pattern recognition control. In non- EMG MLLP, the most used are heuristic rule-based classifiers, like FSM impedance controllers [116] or decision tree controllers [142]. By providing the right set of rules, these controllers can identify a gait phase [115] and additionally identify different types of locomotion, such as walking or slope climbing, driving the mid-level impedance control accordingly [117, 118]. The same purpose of heuristic rule-based classifiers can be achieved with automated pattern recognition algorithms through ML. Such classifiers are usually more robust to outliers but they all require a priori offline training, often on data derived directly from the final user. Linear discriminant analysis (LDA) [140], quadratic discriminant analysis (QDA), Gaussian Mixture Models (GMM) [128], Support Vector Machines (SVM) [66] and Artificial Neural Networks (ANN) [5] are examples of possible solutions that have been explored, each with relative merits and drawbacks. Due to these properties, the ML approaches are usually preferred for EMG-driven pattern recognition controllers on neural signals features [32, 53, 65, 66, 86, 120].

This review has not found any use of Deep Learning (DL) applications for EMG-driven control in MLLPs, in contrast to their upper limb counterparts. Note that DL is a subset of ML, characterized by the ability to self-learn meaningful signal characteristics, rather than requiring features to be defined in advance. The choice of features for classification is a challenging process and impacts long-term performance of the EMG-driven controller [125]. For this reason DL classifiers might represent a solution for future implementation aiming in improving state-of-the-art ML performances [22].

In addition, this review shows that ML strategies are not limited to CIC: they can be used as mapping tool for trajectory generations in IECs, such as in [24, 37], or for reducing signals variability [56, 57].

The diversification regarding the type of neuro-control found in literature complicates the comparison of all studies: the presence of multiple control and processing methods makes it difficult to isolate the effects of each choice on the final outcome. Instead, as presented in the following section, the performance validation metrics and their analysis highly depend on the control design choices.

Performance assessment

The first attempts to introduce EMG signals in the control loop of lower limb prostheses were made in the 70’s with the work of Horn [64] and Donath [29]. Although myoelectric controllers have evolved in the last 40 years, current solutions are still far from being integrated into lower limb prosthetic commercial devices. The main obstacle to this is represented by safety and risk assessment after EMG is introduced in the control loop.

Consistently with the available literature on myoelectric controllers, the self-reported major limitations of EMG-driven controllers for MLLPs are related to instabilities generated from high sensitivities to EMG noise, movement artifacts, and electrode impedance, usually due to skin perspiration. Additionally, pattern recognition approaches mainly report limitations about the fact that only a limited number of locomotion types or motor tasks can be controlled, as already discussed in “EMG-driven working principles” section. The main impediment to model-based control frameworks is represented by the manual tuning of parameters during experimentation, which is crucial for acceptable control. Instead, in direct control, only one manuscript reported that extensive subject training was necessary to compensate for the increased cognitive load [62]. In fact, users required constant awareness about the prosthesis configuration, since the extension and flexion of the knee joint were completely volitional.

Aside from the specific control implementations, the ultimate purpose of a prosthesis is to operate synergistically with the human body and replicate a behaviour that is as similar as possible to the missing biological limb. Therefore, evaluating the ability to generate human-like joint trajectories and forces during locomotion represents an important indication of the performance of the control. Additionally, investigation of performance-based measures, such as changes in cadence, stride length, walking speed, and their application to standard prosthesis controllers represents a strong validation of the performance with respect to the state-of-the-art.

As reported in “Performance assessment” section, these two types of investigations were usually not conducted simultaneously. Depending on the type of neuro-control, either performance-based or biomechanical measurements are used for performance analysis. This particular finding makes it difficult for this review to assess a generic comparison between controllers based on their reported findings, since no common accuracy and reliability metrics could be found.

An additional challenge was found when attempting to compare MLLPs performances with able-bodied kinematics and dynamics, because the results were often qualitative or limited in terms of statistical analysis. [76, 106, 136]. CIC controllers—while usually providing more accurate quantitative analyses with respect to IEC—are usually only limited to classification accuracy of motor task and the type of locomotion recognition. Although it is a solid parameter for the evaluation of FSM controllers, it does not provide a clear measure of the system capability. Few works instead, compared the accuracy of their EMG-driven approach with respect to standard or commercially available CIC [59, 113]. Only few other manuscripts proposed analytical correlations between (1) the recorded accuracy, (2) the final performance of the controller, and (3) the errors in the classification process that could lead to a dangerous failure [59, 145].

Additionally, related to this problem, this review has identified critical lack of attention to significant measurements, such as the control delay. While this parameter is not a direct measurement of the high-level control, it is of importance for the evaluation of the whole HRI.

These issues refer to the concept of bench-marking: the definition of guidelines to be followed in the iterative process of design and development of a new generation of MLLPs. While accredited guidelines are available for testing exoskeletons [91] and bipedal robots [122], standardized experimental methods for evaluating MLLPs are lacking [48]. With this purpose in mind, recent reviews have categorized the evaluation on MLLPs in three methodologies: patient-reported outcomes, performance-based measures and biomechanical measures [42, 48]. Moreover, these reviews highlight the discrepancy between the evaluation metrics used for research and commercial MLLPs, which focus instead on questionnaires on comfort, sense of security and quality of life [101]. This trend can be explained due to the fact that the commercial market is more focused on end-user satisfaction for marketing purposes, whereas researchers tend to focus on engineering performance metrics.

Before arriving to final conclusions, it is necessary to contextualize the obtained results in the clinical situation of lower limb amputations. This discussion is critical to understand the real necessities in this health sector and how myoelectric MLLPs can constructively collaborate to address them.

Clinical considerations

The literature depicts a particularly serious perspective for lower limb amputations and prosthesis utilization. Despite the improvement of medical treatments, the lower limb amputation rate has not substantially changed [38, 105]. This is due to the fact that the population at-risk for lower limb amputation (elderly and diabetic patients) is constantly increasing. In fact, vascular related diseases remain the leading cause for lower limb amputations in industrialized countries [28, 49, 90, 121].

Considering clinical aspects, it is important to note that critical psychosomatic conditions and secondary injuries can emerge over time in amputees. Indeed, evidence suggests that both phantom limb pain and joint degeneration due to compensatory movements can be avoided with neural interfaces and powered prostheses [78, 96, 134].

Moreover, relatively high prosthesis abandonment is influenced by the acceptance rate of lower limb prostheses, which are associated with user perception of inadequate controllability of the device, specifically a lack of intuitiveness in the control [4]. In addition to deficiencies that result from aforementioned mechanical challenges (e.g. noise, heaviness, battery life), the lack of intuitive control likely contributes to the relatively low popularity of powered prostheses. This aspect advocates for the discrepancy between research and the commercial prostheses reported in “Performance assessment” section, where patients’ reports play a major role with respect to the performance and biomechanical capability of the device.

Therefore, implementing volitional control can potentially result in large improvements in the usability of these prostheses. In fact, MLLPs can already recognize and control most of the locomotion modes. However, important classes of motor tasks involving the knee and ankle have been omitted from this set of movements, making it difficult or impossible for amputees to perform them [56]. For example, body shifting position while sitting or particular joint flexion and extension motions are essential during daily activities, such as donning and doffing a shoe, or entering a car. These movements can be executed only using a prosthesis with neural-driven control, unless the repositioning of the joints must be done by hand. Thus, the implementation of this class of movements and the increase of the number of activities in powered prostheses would augment user freedom of motion in daily life. This is the key in making the prosthesis become part of the user routine, which will increase the feeling of embodiment, and will lower the probability to abandon the prosthesis [13].

However, the results of this review have shown several drawbacks (e.g., subject-specificity and high noise sensitivity of EMG signals) related to the adoption of EMG-driven control for MLLPs and no solid representative emerged as the predominant solution. Therefore, the adoption of such technology still remains disputable for MLLPs. Nonetheless, it is important to consider that novel invasive technologies grant a direct connection with the neuromusculoskeletal system (e.g., osteointegration, TMR, implantable EMG sensors). These techniques have been clinically tested in humans and demonstrated the potential to attenuate such limitations of standard myoelectric MLLPs [36]. In summary, it should be possible to substantially improve the performance of neuro-controlled MLLPs by creating a more reliable connection between the robotic limb and the human.

Conclusions

This review has assessed the current state-of-the-art in EMG-driven control methods for MLLPs in the attempt to identify their prospects and limitations, and to formulate suggestions on future research and development. Four major topics of investigation were addressed: neuro-control of the device, EMG-driven working principles, neural input and processing, and performance assessment.

The authors found an evident lack of quantitative and standardized measures regarding sensitivity and risk analysis. Additionally, this review has not found evidence of meaningful comparisons of EMG-driven control in MLLPs with respect to standard adaptive control. Likewise, no comparison was found with commercial products. Moreover, due to missing guidelines in MLLPs development and evaluation, performances metrics are tailored to the type of the implemented neuro-controller. This issue complicates the execution of a quantitative analysis between the reviewed EMG-driven controllers.

However, the reviewed literature also suggests that there are certain preferences regarding control principles depending on the type of movement intended to be restored. If the interest is to provide the prosthesis control during rhythmical locomotion, pattern recognition controllers are to be preferred. This solution is suggested for non-active people that use the prosthesis only for simple daily tasks (e.g., walking, sitting, ascending and descending stairs). For these tasks, this type of controller guarantees a lower cognitive load and more stability. In contrast, control of volitional movements is preferred with direct and model-based controllers. This approach lets amputees control their prosthesis during complex tasks with higher freedom and autonomy, at the expense of higher cognitive effort and attention.

Despite the increasing interest in EMG-driven controllers for MLLPs, a reliable and effective approach that can be introduced into powered prostheses has not yet been formulated. Particularly, continuous developments in ML have resulted in a persisting focus on pattern recognition techniques, which despite allowing for high stability and easier integration, remain a risky approach as they have not proven to be able to tackle the inadequate user control yet. In addition, the full potential of using neural signals is left unexploited, such as targeting direct and voluntary manipulation of the prosthetic joint, which would give the user complete freedom of movement. Based on the evidence reported here, the authors believe that introducing a reliable and effective control, able to integrate both rhythmic (CIC) and volitional (IEC) motor tasks, will promote the use of powered MLLPs despite their current limitations. First attempts in this direction have been investigated in this review. They have demonstrated the possibility to exploit both the potentialities of the two neuro-controllers.

The real efficacy of EMG-driven controllers have yet to be clinically validated. Furthermore, EMG-driven controllers still have to overcome inherent drawbacks such as high noise sensitivity of EMG signals, elevated inter- and intra- subject signals variability, and a difficult integration of electrodes in the socket. However, current advancements in invasive solutions for bionic prostheses have demonstrated to strengthen the connection of the HRI, creating a more reliable neuro-control and attenuating the aforementioned limitations. Therefore, it is likely that the combination of novel invasive interfaces, more advanced decoding algorithms (e.g., DL or modeling) and seamless EMG-driven control will eventually promote the use of powered MLLPs. This claim is strongly related to clinical evidences about the necessity to provide amputees with powered support to avoid psychosomatic conditions and secondary degenerative musculoskeletal pathologies.

Eventually, neural-driven controllers will acquire the capability of closing the sensori-motor loop. Users will be provided with sensory information about touch and proprioception, through the prosthesis, to modulate the control in closed-loop, allowing bioengineers to get closer to the goal of giving back what amputees have lost.

Availability of data and materials

Not applicable.

Abbreviations

ANN:

Artificial Neural Networks

CIC:

Computational Intrinsic Control

CNS:

Central Nervous System

CPGs:

Central-Pattern Generators

DL:

Deep Learning

EMG:

Electromyography

FSM:

Finite-state machine

GMM:

Gaussian Mixture Models

HITL:

Human-in-the-loop

HRI:

Human–Robot Interaction

IEC:

Interactive Extrinsic Control

IMU:

Inertial Measurement Units

LDA:

Linear discriminant analysis

ML:

Machine learning

MLLPs:

Microprocessor-controlled lower limb prostheses

MS:

Musculoskeletal

QDA:

Quadratic discriminant analysis

SlA:

Slope adaptation

SpA:

Speed adaptation

SVM:

Support Vector Machines

TMR:

Targeted Muscle Reinnervation

References

  1. Ahsan R, Ibrahimy MI. EMG signal classification for human computer interaction: a review. Eur J Sci Res. 2009;33(3):480–501.

    Google Scholar 

  2. Aoi S, Ogihara N, Funato T, Sugimoto Y, Tsuchiya K. Evaluating functional roles of phase resetting in generation of adaptive human bipedal walking with a physiologically based model of the spinal pattern generator. Biol Cybern. 2010;102:373–87.

    Article  PubMed  Google Scholar 

  3. Asghari Oskoei M, Hu H. Myoelectric control systems—a survey. Biomed Signal Process Control. 2007;2(4):275–94.

    Article  Google Scholar 

  4. Atkins DJ, Donovan WH. Retrospective analysis of 87 children and adults fitted with electric prosthetic componentry. Arch Phys Med Rehabil. 1992;73(10):960.

    Google Scholar 

  5. Au S, Berniker M, Herr H. Powered ankle-foot prosthesis to assist level-ground and stair-descent gaits. Neural Netw J. 2008;21:654–66.

    Article  Google Scholar 

  6. Au SK, Bonato P, Herr H. An EMG-position controlled system for an active ankle-foot prosthesis: an initial experimental study. In: 9th international conference on rehabilitation robotics, 2005. ICORR 2005, Chicago, IL, USA. IEEE; 2005. p. 375–9.

  7. Au SK, Dilworth P, Herr H. An ankle-foot emulation system for the study of human walking biomechanics. In: Proceedings of the IEEE international conference on robotics and automation, Orlando, FL, USA. IEEE; 2006. p. 2939–45.

  8. Behr J, Friedly J, Molton I, Morgenroth D, Jensen MP, Smith DG. Pain and pain-related interference in adults with lower-limb amputation: comparison of knee-disarticulation, transtibial, and transfemoral surgical sites. J Rehabil Res Dev. 2009;46(7):963–72.

    Article  PubMed  PubMed Central  Google Scholar 

  9. Biddiss E, Chau T. Upper-limb prosthetics: critical factors in device abandonment. Am J Phys Med Rehabilit. 2007;86(12):977–87.

    Article  Google Scholar 

  10. Bohannon RW. Number of pedometer-assessed steps taken per day by adults: a descriptive meta-analysis. Phys Ther. 2007;87(12):1642–50.

    Article  PubMed  Google Scholar 

  11. Brantley JA, Luu TP, Nakagome S, Contreras-Vidal JL. Prediction of lower-limb joint kinematics from surface EMG during overground locomotion. In: IEEE, editor. 2017 IEEE international conference on systems, man, and cybernetics, SMC 2017, Banff, Canada. 2017. p. 1705–9.

  12. Burke MJ, Roman V, Wright V. Bone and joint changes in lower limb amputees. Ann Rheum Dis. 1978;37(3):252–4.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  13. Burrough SF, Brook JA. Patterns of acceptance and rejection of upper limb prostheses. Orthot Prosthet. 1985;39(2):40–7.

    Google Scholar 

  14. Canino JM, Fite KB. Haptic feedback in lower-limb prosthesis: combined haptic feedback and EMG control of a powered prosthesis. In: 2016 IEEE EMBS international student conference (ISC), Ottawa, ON, Canada. IEEE; 2016. p. 1–4.

  15. Cao Y, Gao F, Yu L, She Q. Gait recognition based on emg information with multiple features. IFIP Adv Inf Commun Technol. 2018;538:402–11.

    Article  Google Scholar 

  16. Chen B, Wang Q, Wang L. Promise of using surface EMG signals to volitionally control ankle joint position for powered transtibial prostheses. In: 2014 36th annual international conference of the IEEE engineering in medicine and biology society, EMBC 2014. 2014. p. 2545–8.

  17. Chen B, Wang Q, Wang L. Adaptive slope walking with a robotic transtibial prosthesis based on volitional EMG control. IEEE/ASME Trans Mechatron. 2015;20(5):2146–57.

    Article  Google Scholar 

  18. Ciaccio EJ, Akay M, Dunn SM. Biosignal pattern recognition and interpretation systems. IEEE Eng Med Biol Mag. 1993;12(3):89–95.

    Article  Google Scholar 

  19. Cimolato A, Milandri G, Mattos LS, De Momi E, Laffranchi M, De Michieli L. Hybrid machine learning-neuromusculoskeletal modeling for control of lower limb prosthetics. In: Proceedings of the IEEE RAS and EMBS international conference on biomedical robotics and biomechatronics, vol. 2020-Novem. 2020. p. 557–63.

  20. Cohen AH, Boothe DL. Sensorimotor interactions during locomotion: principles derived from biological systems. Auton Robot. 1999;7(3):239–45.

    Article  Google Scholar 

  21. Cordella F, Ciancio AL, Sacchetti R, Davalli A, Cutti AG, Guglielmelli E, Zollo L. Literature review on needs of upper limb prosthesis users. Front Neurosci. 2016;10(209):1–14.

    Google Scholar 

  22. Côté-Allard U, Campbell E, Phinyomark A, Laviolette F, Gosselin B, Scheme E. Interpreting deep learning features for myoelectric control: a comparison with handcrafted features. Front Bioeng Biotechnol. 2020;8(March):1–22.

    Google Scholar 

  23. Dawley JA, Fite KB, Fulk GD. EMG control of a bionic knee prosthesis: exploiting muscle co-contractions for improved locomotor function. In: 2013 IEEE international conference on rehabilitation robotics (ICORR). 2013. p. 1–6.

  24. Delis AL, Carvalho JL, Da Rocha AF, Ferreira RU, Rodrigues SS, Borges GA. Estimation of the knee joint angle from surface electromyographic signals for active control of leg prostheses. Physiol Meas. 2009;30(9):931–46.

    Article  PubMed  Google Scholar 

  25. Delis AL, De Carvalho JLA, Borges GA, De Rodrigues SS, Dos Santos I, Da Rocha AF. Fusion of electromyographic signals with proprioceptive sensor data in myoelectric pattern recognition for control of active transfemoral leg prostheses. In: Proceedings of the 31st annual international conference of the IEEE engineering in medicine and biology society: engineering the future of biomedicine, EMBC 2009. 2009. p. 4755–8.

  26. Delis AL, De Carvalho JLA, Da Rocha AF, De Oliveira Nascimento FA, Borges GA. Knee angle estimation algorithm for myoelectric control of active transfemoral prostheses. Commun Comput Inf Sci. 2010;52:124–35.

    Google Scholar 

  27. Delis AL, De Carvalho JLA, Seisdedos CV, Borges GA, Da Rocha AF. Myoelectric control algorithms for leg prostheses based on data fusion with proprioceptive sensors. In: Proceedings ISSNIP biosignals and biorobotics conference, January. 2010. p. 137–42.

  28. Dillingham TR, Pezzin LE, MacKenzie EJ. Limb amputation and limb deficiency. South Med J. 2002;95(8):875–83.

    PubMed  Google Scholar 

  29. Donath M. Proportional EMG control for above knee prostheses. Ph.D. thesis, Massachusetts Institute of Technology; 1974.

  30. Drew T, Kalaska J, Krouchev N. Muscle synergies during locomotion in the cat: a model for motor cortex control. J Physiol. 2008;586(5):1239–45.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  31. Drew T, Marigold DS. Taking the next step: cortical contributions to the control of locomotion. Curr Opin Neurobiol. 2015;33:25–33.

    Article  CAS  PubMed  Google Scholar 

  32. Du L, Zhang F, He H, Huang H. Improving the performance of a neural-machine interface for prosthetic legs using adaptive pattern classifiers. In: 2013 35th annual international conference of the IEEE engineering in medicine and biology society (EMBC), Osaka, Japan. IEEE; 2013. p. 1571–4.

  33. Du L, Zhang F, Liu M, Huang H. Toward design of an environment-aware adaptive locomotion-mode-recognition system. IEEE Trans Biomed Eng. 2012;59(10):2716–25.

    Article  PubMed  PubMed Central  Google Scholar 

  34. Eilenberg MF, Geyer H, Herr H. Control of a powered ankle-foot prosthesis based on a neuromuscular model. IEEE Trans Neural Syst Rehabil Eng. 2010;18(2):164–73.

    Article  PubMed  Google Scholar 

  35. Englehart K, Hudgins B. A robust, real-time control scheme for multifunction myoelectric control. IEEE Trans Biomed Eng. 2003;50(7):848–54.

    Article  PubMed  Google Scholar 

  36. Farina D, Vujaklija I, Brånemark R, Bull AM, Dietl H, Graimann B, Hargrove LJ, Hoffmann K-P, Huang HH, Ingvarsson T, et al. Toward higher-performance bionic limbs for wider clinical use. Nat Biomed Eng. 2021. https://doi.org/10.1038/s41551-021-00732-x.

    Article  PubMed  Google Scholar 

  37. Farmer S, Silver-Thorn B, Voglewede P, Beardsley SA. Within-socket myoelectric prediction of continuous ankle kinematics for control of a powered transtibial prosthesis. J Neural Eng. 2014;11(5):1–8.

    Article  Google Scholar 

  38. Feinglass J, Brown JL, LoSasso A, Sohn MW, Manheim LM, Shah SJ, Pearce WH. Rates of lower-extremity amputation and arterial reconstruction in the United States, 1979 to 1996. Am J Public Health. 1999;89(8):1222–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  39. Fleming A, Huang S, Huang H. Proportional myoelectric control of a virtual inverted pendulum using residual antagonistic muscles: toward voluntary postural control. IEEE Trans Neural Syst Rehabil Eng. 2019;27(7):1473–82.

    Article  PubMed  Google Scholar 

  40. Fleming A, Huang S, Huang HH. Coordination of voluntary residual muscle contractions in transtibial amputees: a pilot study. In: Proceedings of the annual international conference of the IEEE engineering in medicine and biology society, EMBS, 2018-July. 2018. p. 2128–31.

  41. Fleming A, Stafford N, Huang S, Hu X, Ferris DP, Huang HH. Myoelectric control of robotic lower limb prostheses: a review of electromyography interfaces, control paradigms, challenges and future directions. J Neural Eng. 2021. https://doi.org/10.1088/1741-2552/ac1176.

    Article  PubMed  PubMed Central  Google Scholar 

  42. Fluit R, Prinsen E, Wang S, Van Der Kooij H. A comparison of control strategies in commercial and research knee prostheses. IEEE Trans Biomed Eng. 2019;67(1):277–90.

    Article  PubMed  Google Scholar 

  43. Frossard L, Laux S, Geada M, Heym PP, Lechler K. Load applied on osseointegrated implant by transfemoral bone-anchored prostheses fitted with state-of-the-art prosthetic components. Clin Biomech. 2021;89: 105457.

    Article  Google Scholar 

  44. Gail J. Using administrative healthcare records to identify determinants of amputee residuum outcomes. Doctoral, Walden University; 2017.

  45. Gailey R, Allen K, Castles J, Kucharik J, Roeder M. Review of secondary physical conditions associated with lower-limb amputation and long-term prosthesis use. J Rehabil Res Dev. 2008;45(1):15–29.

    Article  PubMed  Google Scholar 

  46. Gailey R, McFarland LV, Cooper RA, Czerniecki J, Gambel JM, Hubbard S, Maynard C, Smith DG, Raya M, Reiber GE. Unilateral lower-limb loss: prosthetic device use and functional outcomes in servicemembers from Vietnam war and OIF/OEF conflicts. J Rehabilit Res Dev. 2010;47(4):317.

    Article  Google Scholar 

  47. Geethanjali P. Myoelectric control of prosthetic hands: state-of-the-art review. Med Devices Evid Res. 2016;9:247–55.

    Article  Google Scholar 

  48. Ghillebert J, De Bock S, Flynn L, Geeroms J, Tassignon B, Roelands B, Lefeber D, Vanderborght B, Meeusen R, De Pauw K. Guidelines and recommendations to investigate the efficacy of a lower-limb prosthetic device: a systematic review. IEEE Trans Med Robot Bionics. 2019;1(4):279–96.

    Article  Google Scholar 

  49. Godlwana L, Nadasan T, Puckree T. Global trends in incidence of lower limb amputation: a review of the literature. S Afr J Physiother. 2008;64(1):8–12.

    Article  Google Scholar 

  50. Goldfarb M, Lawson BE, Shultz AH. Realizing the promise of robotic leg prostheses. Sci Transl Med. 2013;5(210):1–5.

    Article  Google Scholar 

  51. Gottschalk F. Transfemoral amputation. Clin Orthop Relat Res. 1999;361(361):15–22.

    Article  Google Scholar 

  52. Grimmer M, Seyfarth A. Chapter 5: Mimicking human-like leg function in prosthetic limbs. In: Artemiadis P, editor. Neuro-robotics, vol. 2. Trends in augmentation of human performance. Darmstadt: Springer Science+Business Media; 2014. p. 105–55.

    Google Scholar 

  53. Guo X, Chen L, Zhang Y, Yang P, Zhang L. A study on control mechanism of above knee robotic prosthesis based on CPG model. In: 2010 IEEE international conference on robotics and biomimetics (ROBIO), Tianjin, China. IEEE; 2010. p. 283–7.

  54. Gupta R, Agarwal R. Continuous human locomotion identification for lower limb prosthesis control. CSI Trans ICT. 2018;6(1):17–31.

    Article  Google Scholar 

  55. Gupta R, Agarwal R. Single channel EMG-based continuous terrain identification with simple classifier for lower limb prosthesis. Biocybern Biomed Eng. 2019;39(3):775–88.

    Article  Google Scholar 

  56. Ha KH, Varol HA, Goldfarb M. Volitional control of a prosthetic knee using surface electromyography. IEEE Trans Biomed Eng. 2011;58(1):144–51.

    Article  PubMed  Google Scholar 

  57. Hargrove LJ, Simon AM, Lipschutz R, Finucane SB, Kuiken TA. Non-weight-bearing neural control of a powered transfemoral prosthesis. J Neuroeng Rehabil. 2013;10(1):62.

    Article  PubMed  PubMed Central  Google Scholar 

  58. Hargrove LJ, Simon AM, Young AJ, Lipschutz RD, Finucane SB, Smith DG, Kuiken TA. Robotic leg control with EMG decoding in an amputee with nerve transfers. N Engl J Med. 2013;369(13):1237–42.

    Article  CAS  PubMed  Google Scholar 

  59. Hargrove LJ, Young AJ, Simon AM, Fey NP, Lipschutz RD, Finucane SB, Halsne EG, Ingraham KA, Kuiken TA. Intuitive control of a powered prosthetic leg during ambulation: a randomized clinical trial. JAMA. 2015;313(22):2244–52.

    Article  CAS  PubMed  Google Scholar 

  60. Hof AL, van Bockel RM, Schoppen T, Postema K. Control of lateral balance in walking. Experimental findings in normal subjects and above-knee amputees. Gait Posture. 2007;25(2):250–8.

    Article  PubMed  Google Scholar 

  61. Hoover CD, Fite KB. A configuration dependent muscle model for the myoelectric control of a transfemoral prosthesis. In: 2011 IEEE international conference on rehabilitation robotics (ICORR), Zurich, Switzerland. IEEE; 2011. p. 1–6.

  62. Hoover CD, Fulk GD, Fite KB. Stair ascent with a powered transfemoral prosthesis under direct myoelectric control. IEEE/ASME Trans Mechatron. 2012;18(3):1191–200.

    Article  Google Scholar 

  63. Hoover CD, Fulk GD, Fite KB. The design and initial experimental validation of an active myoelectric transfemoral prosthesis. J Med Devices. 2012;6(1): 011005.

    Article  Google Scholar 

  64. Horn GW. Electro-control: am EMG-controlled A/K prosthesis. Med Biol Eng. 1972;10(1):61–73.

    Article  CAS  PubMed  Google Scholar 

  65. Huang H, Kuiken TA, Member S, Lipschutz RD. A strategy for identifying locomotion modes using surface electromyography. IEEE Trans Biomed Eng. 2009;56(1):65–73.

    Article  PubMed  PubMed Central  Google Scholar 

  66. Huang H, Zhang F, Hargrove LJ, Dou Z, Rogers DR, Englehart KB. Continuous locomotion-mode identification for prosthetic legs based on neuromuscular—mechanical fusion. IEEE Trans Biomed Eng. 2011;58(10):2867–75.

    Article  PubMed  PubMed Central  Google Scholar 

  67. Huang S, Wensman JP, Ferris DP. An experimental powered lower limb prosthesis using proportional myoelectric control. J Med Devices Trans ASME. 2014;8(2):1–5.

    Google Scholar 

  68. Huang S, Wensman JP, Ferris DP. Locomotor adaptation by transtibial amputees walking with an experimental powered prosthesis under continuous myoelectric control. IEEE Trans Neural Syst Rehabil Eng. 2016;24(5):573–81.

    Article  PubMed  Google Scholar 

  69. Jiménez-Fabián R, Verlinden O. Review of control algorithms for robotic ankle systems in lower-limb orthoses, prostheses, and exoskeletons. Med Eng Phys. 2012;34(4):397–408.

    Article  PubMed  Google Scholar 

  70. Jin D, Zhang R, Zhang J, Wang R, Gruver W. An intelligent above-knee prosthesis with EMG-based terrain Identification. In: 2000 IEEE international conference on systems, man, and cybernetics, vol. 3, Nashville, TN, USA. IEEE; 2000. p. 1859–64.

  71. Kannape OA, Herr HM. Volitional control of ankle plantar flexion in a powered transtibial prosthesis during stair-ambulation. In: 36th annual international conference of the IEEE engineering in medicine and biology society, Chicago, IL, USA. IEEE; 2014. p. 1662–5.

  72. Kannape OA, Herr HM. Split-belt adaptation and gait symmetry in transtibial amputees walking with a hybrid EMG controlled ankle-foot prosthesis. In: 2016 38th annual international conference of the IEEE engineering in medicine and biology society (EMBC), Orlando, FL, USA. IEEE; 2016. p. 5469–72.

  73. Kawato M, Furukawa K, Suzuki R. A hierarchical neural-network model for control and learning of voluntary movement. Biol Cybern. 1987;57(3):169–85.

    Article  CAS  PubMed  Google Scholar 

  74. Konrad P. The ABC of EMG, vol. 1. 1.4. Scottsdale: Noxar INC.; 2005.

    Google Scholar 

  75. Kusljugic A, Kapidzic-Durakovic S, Kudumovic Z, Cickusic A. Chronic low back pain in individuals with lower-limb amputation. Bosn J Basic Med Sci. 2006;6(2):67–70.

    Article  PubMed  PubMed Central  Google Scholar 

  76. Jacob T, Saini LM, Bhaumick S. An algorithm for control of prosthetic foot by gait characteristics. In: Proceeding in international conference on energy, communication, data analytics and soft computing. IEEE; 2017. p. 3552–6.

  77. Latham PE, Nirenberg S. Synergy, redundancy, and independence in population codes, revisited. J Neurosci. 2005;25(21):5195–206.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  78. Lendaro E, Mastinu E, Håkansson B, Ortiz-Catalan M. Real-time classification of non-weight bearing lower-limb movements using EMG to facilitate phantom motor execution: engineering and case study application on phantom limb pain. Front Neurol. 2017;8(SEP):1–12.

    Google Scholar 

  79. Li W, Sadigh D, Shankar Sastry S, Seshia SA. Synthesis for human-in-the-loop control systems. In: Erika Á, Havelund K, editors. International conference on tools and algorithms for the construction and analysis of systems. Berlin: Springer; 2014. p. 470–84.

    Google Scholar 

  80. Lotze M, Grodd W, Birbaumer N, Erb M, Huse E, Flor H. Does use of a myoelectric prosthesis prevent cortical reorganization and phantom limb pain? Nat Neurosci. 1999;2(6):501–2.

    Article  CAS  PubMed  Google Scholar 

  81. MacKay-Lyons M. Central pattern generation of locomotion: a review of the evidence. Phys Ther. 2002;82(1):69–83.

    Article  PubMed  Google Scholar 

  82. Markin SN, Klishko AN, Shevtsova NA, Lemay MA, Prilutsky BI, Rybak IA. Afferent control of locomotor CPG: insights from a simple neuromechanical model. Ann N Y Acad Sci. 2010;1198(1):21–34.

    Article  PubMed  PubMed Central  Google Scholar 

  83. Markowitz J, Krishnaswamy P, Eilenberg MF, Endo K, Barnhart C, Herr H. Speed adaptation in a powered transtibial prosthesis controlled with a neuromuscular model. Philos Trans R Soc B Biol Sci. 2011;366(1570):1621–31.

    Article  Google Scholar 

  84. Martin J, Pollock A, Hettinger J. Microprocessor lower limb prosthetics: review of current state of the art. J Orthot Prosthet. 2010;22(3):183–93.

    Article  Google Scholar 

  85. Meng M, Luo Z, She Q, Ma Y. Automatic recognition of gait mode from EMG signals of lower limb. In: ICIMA 2010–2010 2nd international conference on industrial mechatronics and automation, vo. 1. 2010. p. 282–5.

  86. Miller JD, Beazer MS, Hahn ME. Myoelectric walking mode classification for transtibial amputees. IEEE Trans Biomed Eng. 2013;60(10):2745–50.

    Article  PubMed  Google Scholar 

  87. Miller JD, Seyedali M, Hahn ME. Walking mode classification from myoelectric and inertial fusion. In: ASME, editor. Proceedings of ASME 2012 summer bioengineering conference, Fajardo, Puerto Rico. 2012. p. 2–3.

  88. Miller WC, Deathe AB, Speechley M, Koval J. The influence of falling, fear of falling, and balance confidence on prosthetic mobility and social activity among individuals with a lower extremity amputation. Arch Phys Med Rehabil. 2001;82(9):1238–44.

    Article  CAS  PubMed  Google Scholar 

  89. Miller WC, Speechley M, Deathe AB. Balance confidence among people with lower-limb amputations. Phys Ther. 2002;82(9):856–65.

    Article  PubMed  Google Scholar 

  90. Moxey PW, Gogalniceanu P, Hinchliffe RJ, Loftus IM, Jones KJ, Thompson MM, Holt PJ. Lower extremity amputations—a review of global variability in incidence. Diabet Med. 2011;28(10):1144–53.

    Article  CAS  PubMed  Google Scholar 

  91. Mudie KL, Boynton AC, Karakolis T, O’Donovan MP, Kanagaki GB, Crowell HP, Begg RK, LaFiandra ME, Billing DC. Consensus paper on testing and evaluation of military exoskeletons for the dismounted combatant. J Sci Med Sport. 2018;21(11):1154–61.

    Article  PubMed  Google Scholar 

  92. Naschitz JE, Lenger R. Why traumatic leg amputees are at increased risk for cardiovascular diseases. Q J Med. 2008;101(4):251–9.

    Article  CAS  Google Scholar 

  93. Navarro X, Krueger TB, Lago N, Micera S, Stieglitz T, Dario P. A critical review of interfaces with the peripheral nervous system for the control of neuroprostheses and hybrid bionic systems. J Peripher Nerv Syst. 2005;10(3):229–58.

    Article  PubMed  Google Scholar 

  94. Nazmi N, Abdul Rahman M, Yamamoto S-I, Ahmad S, Zamzuri H, Mazlan S. A review of classification techniques of EMG signals during isotonic and isometric contractions. Sensors. 2016;16(1304):1–28.

    Google Scholar 

  95. Nolan L, Wit A, Dudziñski K, Lees A, Lake M, Wychowañski M. Adjustments in gait symmetry with walking speed in trans-femoral and trans-tibial amputees. Gait Posture. 2003;17(2):142–51.

    Article  PubMed  Google Scholar 

  96. Ortiz-Catalan M, Sander N, Kristoffersen MB, Håkansson B, Brånemark R. Treatment of phantom limb pain (PLP) based on augmented reality and gaming controlled by myoelectric pattern recognition: a case study of a chronic PLP patient. Front Neurosci. 2014;8(24):1–7.

    Google Scholar 

  97. Parker P, Englehart K, Hudgins B. Myoelectric signal processing for control of powered limb prostheses. J Electromyogr Kinesiol. 2006;16(6):541–8.

    Article  CAS  PubMed  Google Scholar 

  98. Peeraer L, Aeyels B, Van der Perre G. Development of EMG-based mode and intent recognition algorithms for a computer-controlled above-knee prosthesis. J Biomed Eng. 1990;12(3):178–82.

    Article  CAS  PubMed  Google Scholar 

  99. Pitkin MR. Lower limb prosthesis. In: Biomechanics of lower limb prosthetics, vol. xx. Berlin: Springer; 2010. p. 1–27.

    Chapter  Google Scholar 

  100. Popovic D, Tomovic R, Tepavac D, Schwirtlich L. Control aspects of active above-knee prosthesis. Int J Man Mach Stud. 1991;35(6):751–67.

    Article  Google Scholar 

  101. Radcliffe CW. Functional considerations in the fitting of above-knee prostheses. Artif Limbs. 1955;2(1):35–60.

    CAS  PubMed  Google Scholar 

  102. Resnik L, Huang HH, Winslow A, Crouch DL, Zhang F, Wolk N. Evaluation of EMG pattern recognition for upper limb prosthesis control: a case study in comparison with direct myoelectric control. J Neuroeng Rehabil. 2018;15(1):1–13.

    Article  Google Scholar 

  103. Roche AD, Rehbaum H, Farina D, Aszmann OC. Prosthetic myoelectric control strategies: a clinical perspective. Curr Surg Rep. 2014;2(3):1–11.

    Article  Google Scholar 

  104. Roffman CE, Buchanan J, Allison GT. Predictors of non-use of prostheses by people with lower limb amputation after discharge from rehabilitation: development and validation of clinical prediction rules. J Physiother. 2014;60(4):224–31.

    Article  PubMed  Google Scholar 

  105. Santosa F, Kroger K. Chapter 3: Trends in amputation. In: Vitin A, editor. Gangrene management—new advancements and current trends. Rijeka: IntechOpen; 2013. p. 27–36.

    Google Scholar 

  106. Sawake N, Gupta S, Ghatge A, Khatri A. EMG-based prosthetic leg for above-knee amputee. In: 2014 Texas instruments India educators’ conference (TIIEC), Los Alamitos, CA, US. IEEE; 2016. p. 69–72.

  107. Scheme E, Englehart K. Electromyogram pattern recognition for control of powered upper-limb prostheses: state of the art and challenges for clinical use. J Rehabilit Res Dev. 2011;48(6):643.

    Article  Google Scholar 

  108. Schultz AE, Kuiken TA. Neural interfaces for control of upper limb prostheses: the state of the art and future possibilities. PM&R. 2011;3(1):55–67.

    Article  Google Scholar 

  109. Scott RN, Parker PA. Myoelectric prostheses: state of the art. J Med Eng Technol. 1988;12(4):143–51.

    Article  CAS  PubMed  Google Scholar 

  110. She Q, Luo Z, Meng M, Xu P. Multiple kernel learning SVM-based EMG pattern classification for lower limb control. In: 11th international conference on control, automation, robotics and vision, ICARCV 2010, (December). 2010. p. 2109–13.

  111. Simon AM, Fey NP, Ingraham KA, Young AJ, Hargrove LJ. Powered prosthesis control during walking, sitting, standing, and non-weight bearing activities using neural and mechanical inputs. In: 2013 6th international IEEE/EMBS conference on neural engineering (NER), San Diego, CA, USA. IEEE; 2013. p. 1174–7.

  112. Smail LC, Neal C, Wilkins C, Packham TL. Comfort and function remain key factors in upper limb prosthetic abandonment: findings of a scoping review. Disabil Rehabilit Assist Technol. 2021;16(8):821–30.

    Article  Google Scholar 

  113. Spanias JA, Perreault EJ, Hargrove LJ. Detection of and compensation for EMG disturbances for powered lower limb prosthesis control. IEEE Trans Neural Syst Rehabil Eng. 2016;24(2):226–34.

    Article  PubMed  Google Scholar 

  114. Spanias JA, Simon AM, Ingraham KA, Hargrove LJ. Effect of additional mechanical sensor data on an EMG-based pattern recognition system for a powered leg prosthesis. In: 2015 7th international IEEE/EMBS conference on neural engineering (NER), Montpellier, France. IEEE; 2015. p. 639–42.

  115. Sup F, Bohara A, Goldfarb M. Design and control of a powered knee and ankle prosthesis. In: 2007 IEEE international conference on robotics and automation, Roma, Italy. IEEE; 2007. p. 4134–9.

  116. Sup F, Bohara A, Goldfarb M. Design and control of a powered transfemoral prosthesis. Int J Robot Res. 2008;27(2):263–73.

    Article  Google Scholar 

  117. Sup F, Varol HA, Goldfarb M. Upslope walking with a powered knee and ankle prosthesis: initial results with an amputee subject. IEEE Trans Neural Syst Rehabil Eng. 2011;19(1):71–8.

    Article  PubMed  Google Scholar 

  118. Sup F, Varol HA, Mitchell J, Withrow TJ, Goldfarb M. Preliminary evaluations of a self-contained anthropomorphic transfemoral prosthesis. IEEE/ASME Trans Mechatron. 2009;14(6):667–76.

    Article  Google Scholar 

  119. Suzuki R, Sawada T, Kobayashi N, Hofer EP. Control method for powered ankle prosthesis via internal model control design. In: 2011 international conference on mechatronics and automation (ICMA), Beijing, China. IEEE; 2011. p. 237–42.

  120. Tkach DC, Lipschutz RD, Finucane SB, Hargrove LJ. Myoelectric neural interface enables accurate control of a virtual multiple degree-of-freedom foot-ankle prosthesis. In: IEEE international conference on rehabilitation robotics, Seattle, WA, USA. IEEE; 2013. p. 1–4.

  121. Torrealba RR, Fernández-López G, Grieco JC. Towards the development of knee prostheses: review of current researches. Kybernetes. 2008;37(9–10):1561–76.

    Article  Google Scholar 

  122. Torricelli D, Mizanoor RS, Gonzalez J, Lippi V, Hettich G, Asslaender L, Weckx M, Vanderborght B, Dosen S, Sartori M, Zhao J, Schütz S, Liu Q, Mergner T, Lefeber D, Farina D, Berns K, Pons JL. Benchmarking human-like posture and locomotion of humanoid robots: a preliminary scheme. Lecture notes in computer science (including subseries lecture notes in artificial intelligence and lecture notes in bioinformatics). New York: Springer; 2014. p. 8608.

    Google Scholar 

  123. TroyBlackburn J, Bell DR, Norcross MF, Hudson JD, Engstrom LA. Comparison of hamstring neuromechanical properties between healthy males and females and the influence of musculotendinous stiffness. J Electromyogr Kinesiol. 2009;19(5):e362–9.

    Article  Google Scholar 

  124. Tucker MR, Olivier J, Pagel A, Bleuler H, Bouri M, Lambercy O, Del Millán JR, Riener R, Vallery H, Gassert R. Control strategies for active lower extremity prosthetics and orthotics: a review. J Neuroeng Rehabil. 2015;12(1):1–29.

    Article  PubMed  PubMed Central  Google Scholar 

  125. ur Rehman MZ, Waris A, Gilani SO, Jochumsen M, Niazi IK, Jamil M, Farina D, Kamavuako EN. Multiday EMG-based classification of hand motions with deep learning techniques. Sensors. 2018;18(8):1–16.

    Google Scholar 

  126. Valgeirsdóttir VV, Sigurðardóttir JS, Lechler K, Tronicke L, Jóhannesson ÓI, Alexandersson Á, Kristjánsson Á. How do we measure success? A review of performance evaluations for lower-limb neuroprosthetics. JPO. 2022;34(1):e20–36.

    Google Scholar 

  127. Vallery H, Burgkart R, Hartmann C, Mitternacht J, Riener R, Buss M. Complementary limb motion estimation for the control of active knee prostheses. Biomed Tech. 2011;56(1):45–51.

    Article  Google Scholar 

  128. Varol HA, Sup F, Goldfarb M. Multiclass real-time intent recognition of a powered lower limb prosthesis. IEEE Trans Biomed Eng. 2010;57(3):542–51.

    Article  PubMed  Google Scholar 

  129. Vos EJ, Harlaar J, Schenau GJVI. Electromechanical delay during knee extensor contractions. Med Sci Sport Exerc. 1991;23(10):1187–93.

    Article  CAS  Google Scholar 

  130. Wang J, Kannape OA, Herr HM. Proportional EMG control of ankle plantar flexion in a powered transtibial prosthesis. In: IEEE international conference on rehabilitation robotics. 2013.

  131. Windrich M, Grimmer M, Christ O, Rinderknecht S, Beckerle P. Active lower limb prosthetics: a systematic review of design issues and solutions. Biomed Eng Online. 2016;15(S3):5–19.

    Article  Google Scholar 

  132. Winter D, Yack H. EMG profiles during normal human walking: stride-to-stride and inter-subject variability. Electroencephalogr Clin Neurophysiol. 1987;67(5):402–11.

    Article  CAS  PubMed  Google Scholar 

  133. Winter DA. Kinematic and kinetic patterns in human gait: variability and compensating effects. Hum Mov Sci. 1984;3(1–2):51–76.

    Article  Google Scholar 

  134. Wolf EJ, Everding VQ, Linberg AL, Schnall BL, Czerniecki JM, Gambel JM. Assessment of transfemoral amputees using C-leg and power knee for ascending and descending inclines and steps. J Rehabilit Res Dev. 2012;49(6):831–42.

    Article  Google Scholar 

  135. Wu SK, Waycaster G, Shen X. Active knee prosthesis control with electromyography. In: ASME 2010 dynamic systems and control conference, DSCC2010, vol. 1, Cambridge, Massachusetts, USA. ASME; 2010. p. 785–91.

  136. Wu SK, Waycaster G, Shen X. Electromyography-based control of active above-knee prostheses. Control Eng Pract. 2011;19(8):875–82.

    Article  Google Scholar 

  137. Xiao W, Huang H, Sun Y, Yang Q. Promise of embedded system with GPU in artificial leg control: enabling time-frequency feature extraction from electromyography. In: Proceedings of the 31st annual international conference of the IEEE engineering in medicine and biology society: engineering the future of biomedicine, EMBC 2009. 2009. p. 6926–9.

  138. Yakovenko S, Gritsenko V, Prochazka A. Contribution of stretch reflexes to locomotor control: a modeling study. Biol Cybern. 2004;90(2):146–55.

    Article  CAS  PubMed  Google Scholar 

  139. Young AJ, Kuiken TA, Hargrove LJ. Analysis of using EMG and mechanical sensors to enhance intent recognition in powered lower limb prostheses. J Neural Eng. 2014;11(5): 056021.

    Article  CAS  PubMed  Google Scholar 

  140. Young AJ, Simon AM, Fey NP, Hargrove LJ. Classifying the intent of novel users during human locomotion using powered lower limb prostheses. In: 2013 6th international IEEE/EMBS conference on neural engineering (NER), San Diego, CA, USA. IEEE; 2013. p. 311–4.

  141. Zhang F, Dou Z, Nunnery M, Huang H. Real-time implementation of an intent recognition system for artificial legs. In: Proceedings of the annual international conference of the IEEE engineering in medicine and biology society, EMBS. 2011. p. 2997–3000.

  142. Zhang F, Fang Z, Liu M, Huang H. Preliminary design of a terrain recognition system. In: Proceedings of the annual international conference of the IEEE engineering in medicine and biology society, EMBS, Boston, MA, USA. IEEE; 2011. p. 5452–5.

  143. Zhang F, Huang H. Source selection for real-time user intent recognition toward volitional control of artificial legs. IEEE J Biomed Health Inform. 2013;17(5):907–14.

    Article  PubMed Central  Google Scholar 

  144. Zhang F, Huang HH. Real-time recognition of user intent for neural control of artificial legs. In: Proceedings of the 2011 MyoElectric controls/powered prosthetics symposium fredericton, New Brunswick, Canada. 2011. p. 1–4.

  145. Zhang F, Liu M, Huang H. Preliminary study of the effect of user intent recognition errors on volitional control of powered lower limb prostheses. In: 2012 annual international conference of the IEEE engineering in medicine and biology society (EMBC), San Diego, California, USA. IEEE; 2012. p. 2768–71.

Download references

Acknowledgements

This work was supported by the Istituto Nazionale per l’Assicurazione contro gli Infortuni sul Lavoro (INAIL) under grant agreement PPR-AI.

Funding

Not applicable.

Author information

Authors and Affiliations

Authors

Contributions

AC conceived of and coordinated the reading and writing for this review. He is the main contributor of text and figures; he compiled the data in tabular form. AC reviewed the collected literature and defined four major topics of investigation. JD has contributed in restructuring the document, (re)writing multiple paragraphs and generating graphical overviews. JD helped in the definition of subcategories in each of the four topics to optimize the data collection. JD, LSM, EDM, LDM and ML reviewed the data in the tables and provided critical feedback on all the sections of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Andrea Cimolato.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Cimolato, A., Driessen, J.J.M., Mattos, L.S. et al. EMG-driven control in lower limb prostheses: a topic-based systematic review. J NeuroEngineering Rehabil 19, 43 (2022). https://doi.org/10.1186/s12984-022-01019-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12984-022-01019-1

Keywords